Sensitivity analysis of a computer model of neonatal oxygen transport
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21460%2F20%3A00344324" target="_blank" >RIV/68407700:21460/20:00344324 - isvavai.cz</a>
Výsledek na webu
<a href="https://doi.org/10.1515/cdbme-2020-3026" target="_blank" >https://doi.org/10.1515/cdbme-2020-3026</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1515/cdbme-2020-3026" target="_blank" >10.1515/cdbme-2020-3026</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Sensitivity analysis of a computer model of neonatal oxygen transport
Popis výsledku v původním jazyce
Computer models of neonatal oxygenation could serve as a tool for a comprehensive comparison of closed-loop automated oxygen control systems. The behaviour of such models depends, besides the input data of the inspired fraction of oxygen and the premature infant's breath pattern, on internal parameters of the model. The aim of this study was to perform a sensitivity analysis of a computer model of neonatal oxygen transport to clarify the influence of its internal physiological parameters on the output signal of peripheral oxygen saturation (SpO2). We performed a multi-parameter sensitivity analysis using Monte Carlo simulations for randomly generated values of eight internal parameters. The influence on the model output SpO2 signal was evaluated using five characteristics of the output signal. The relations between the parameters and the output characteristics were displayed using scatter plots and analysed by linear correlation, standardized regression, and partial correlation. The main result of the study is that in our model the oxygen consumption in the tissue and the cardiac output have the greatest influence on the SpO2 drop and minimal SpO2 value during simulated desaturation. The rate of development of desaturation and its duration are most affected by the diffusion resistance of the alveolar-capillary membrane. The results of the sensitivity analysis will help to optimize the performance of the computer model of neonatal oxygen transport.
Název v anglickém jazyce
Sensitivity analysis of a computer model of neonatal oxygen transport
Popis výsledku anglicky
Computer models of neonatal oxygenation could serve as a tool for a comprehensive comparison of closed-loop automated oxygen control systems. The behaviour of such models depends, besides the input data of the inspired fraction of oxygen and the premature infant's breath pattern, on internal parameters of the model. The aim of this study was to perform a sensitivity analysis of a computer model of neonatal oxygen transport to clarify the influence of its internal physiological parameters on the output signal of peripheral oxygen saturation (SpO2). We performed a multi-parameter sensitivity analysis using Monte Carlo simulations for randomly generated values of eight internal parameters. The influence on the model output SpO2 signal was evaluated using five characteristics of the output signal. The relations between the parameters and the output characteristics were displayed using scatter plots and analysed by linear correlation, standardized regression, and partial correlation. The main result of the study is that in our model the oxygen consumption in the tissue and the cardiac output have the greatest influence on the SpO2 drop and minimal SpO2 value during simulated desaturation. The rate of development of desaturation and its duration are most affected by the diffusion resistance of the alveolar-capillary membrane. The results of the sensitivity analysis will help to optimize the performance of the computer model of neonatal oxygen transport.
Klasifikace
Druh
J<sub>SC</sub> - Článek v periodiku v databázi SCOPUS
CEP obor
—
OECD FORD obor
20601 - Medical engineering
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2020
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Current Directions in Biomedical Engineering
ISSN
2364-5504
e-ISSN
2364-5504
Svazek periodika
6
Číslo periodika v rámci svazku
3
Stát vydavatele periodika
DE - Spolková republika Německo
Počet stran výsledku
4
Strana od-do
99-102
Kód UT WoS článku
—
EID výsledku v databázi Scopus
2-s2.0-85097316284