Abstracts of the presentations for the young scientists and professionals award: Fast neutron dosemeter using pixelated detector Timepix
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F86652052%3A_____%2F15%3A%230000332" target="_blank" >RIV/86652052:_____/15:#0000332 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1093/rpd/ncu332" target="_blank" >http://dx.doi.org/10.1093/rpd/ncu332</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/rpd/ncu332" target="_blank" >10.1093/rpd/ncu332</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Abstracts of the presentations for the young scientists and professionals award: Fast neutron dosemeter using pixelated detector Timepix
Popis výsledku v původním jazyce
Possibilities of use of the pixelated detector Timepix covered with a set of polyethylene convertors of different thicknesses for purposes of measuring ambient dose equivalent in fast neutron fields of unknown spectra are presented in the paper. Dose reconstruction is based on evaluation of responses measured for particular convertors. The dosemeter was calibrated for monoenergetic neutrons in the energy of 0.4–16.5 MeV. Results of the pilot experiment proved the potential use of the designed dosemeter for fast neutron dosimetry. Presented at journal "Radiation Protection Dosimetry" as Technical Note.
Název v anglickém jazyce
Abstracts of the presentations for the young scientists and professionals award: Fast neutron dosemeter using pixelated detector Timepix
Popis výsledku anglicky
Possibilities of use of the pixelated detector Timepix covered with a set of polyethylene convertors of different thicknesses for purposes of measuring ambient dose equivalent in fast neutron fields of unknown spectra are presented in the paper. Dose reconstruction is based on evaluation of responses measured for particular convertors. The dosemeter was calibrated for monoenergetic neutrons in the energy of 0.4–16.5 MeV. Results of the pilot experiment proved the potential use of the designed dosemeter for fast neutron dosimetry. Presented at journal "Radiation Protection Dosimetry" as Technical Note.
Klasifikace
Druh
O - Ostatní výsledky
CEP obor
BG - Jaderná, atomová a molekulová fyzika, urychlovače
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/VF20102015014" target="_blank" >VF20102015014: Výzkum pokročilých metod detekce, stanovení a následné zvládnutí radioaktivní kontaminace</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů