A multinational audit of small field output factors calculated by treatment planning systems used in radiotherapy
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F86652052%3A_____%2F18%3AN0000012" target="_blank" >RIV/86652052:_____/18:N0000012 - isvavai.cz</a>
Výsledek na webu
<a href="https://www.phiro.science/article/S2405-6316(17)30048-9/pdf" target="_blank" >https://www.phiro.science/article/S2405-6316(17)30048-9/pdf</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.phro.2018.02.005" target="_blank" >10.1016/j.phro.2018.02.005</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
A multinational audit of small field output factors calculated by treatment planning systems used in radiotherapy
Popis výsledku v původním jazyce
The purpose was to test an audit methodology for verifying the implementation of output factors (OFs) of small fields in treatment planning systems (TPSs) used in radiotherapy. The methodology was developed and tested through a multinational research group and performed on a national level in five different countries. Centres participating in this study were asked to provide OFs calculated by their TPSs for 10×10 cm2, 6×6cm2, 4×4cm2, 3×3cm2 and 2×2cm2 field sizes using an SSD of 100 cm. The ratio of these calculated OFs to reference OFs was analysed. The action limit was ±3% for the 2×2 cm2 field and ±2% for all other fields. OFs for more than 200 different beams were collected in total. On average, the OFs for small fields calculated by TPSs were generally larger than measured reference data. These deviations increased with decreasing field size. On a national level, 30% and 31% of the calculated OFs of the 2×2 cm2 field exceeded the action limit of 3% for nominal beam energies of 6 MV and for nominal beam energies higher than 6 MV, respectively. Modern TPS beam models generally overestimate the OFs for small fields. The verification of calculated small field OFs is a vital step and should be included when commissioning a TPS. The methodology outlined in this study can be used to identify potential discrepancies in clinical beam models.
Název v anglickém jazyce
A multinational audit of small field output factors calculated by treatment planning systems used in radiotherapy
Popis výsledku anglicky
The purpose was to test an audit methodology for verifying the implementation of output factors (OFs) of small fields in treatment planning systems (TPSs) used in radiotherapy. The methodology was developed and tested through a multinational research group and performed on a national level in five different countries. Centres participating in this study were asked to provide OFs calculated by their TPSs for 10×10 cm2, 6×6cm2, 4×4cm2, 3×3cm2 and 2×2cm2 field sizes using an SSD of 100 cm. The ratio of these calculated OFs to reference OFs was analysed. The action limit was ±3% for the 2×2 cm2 field and ±2% for all other fields. OFs for more than 200 different beams were collected in total. On average, the OFs for small fields calculated by TPSs were generally larger than measured reference data. These deviations increased with decreasing field size. On a national level, 30% and 31% of the calculated OFs of the 2×2 cm2 field exceeded the action limit of 3% for nominal beam energies of 6 MV and for nominal beam energies higher than 6 MV, respectively. Modern TPS beam models generally overestimate the OFs for small fields. The verification of calculated small field OFs is a vital step and should be included when commissioning a TPS. The methodology outlined in this study can be used to identify potential discrepancies in clinical beam models.
Klasifikace
Druh
J<sub>ost</sub> - Ostatní články v recenzovaných periodicích
CEP obor
—
OECD FORD obor
20601 - Medical engineering
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2018
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Physics and Imaging in Radiation Oncology
ISSN
2405-6316
e-ISSN
—
Svazek periodika
5
Číslo periodika v rámci svazku
January
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
6
Strana od-do
58-63
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—