ELISA kit for beta-lactoglobulin determination - Collaborative study
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00027022%3A_____%2F09%3A00000038" target="_blank" >RIV/00027022:_____/09:00000038 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
ELISA kit for beta-lactoglobulin determination - Collaborative study
Popis výsledku v původním jazyce
An interlaboratory study was performed in six laboratories to prove the validation of the ELISA method developed for quantitative determination of beta-lactoglobulin (BLG) in foods. The ELISA kit used for this study is based on rabbit polyclonal antibody. In-house validation of the kit did not produce false-positive results or cross-reactivity in a broad range of food matrixes containing no milk proteins. All participants obtained the BLG kit with a standard operational procedure, the list of the samples, samples, and a protocol for recording test results.The statistical tests (Cochran, Dixon, and Mandel) and analysis of variance were used to evaluate the interlaboratory study results. Repeatability and reproducibility limits, as well as LOQ (0.22 mg BLG/kg) and LOD (0.07 mg BLG/kg), for the kit were calculated.
Název v anglickém jazyce
ELISA kit for beta-lactoglobulin determination - Collaborative study
Popis výsledku anglicky
An interlaboratory study was performed in six laboratories to prove the validation of the ELISA method developed for quantitative determination of beta-lactoglobulin (BLG) in foods. The ELISA kit used for this study is based on rabbit polyclonal antibody. In-house validation of the kit did not produce false-positive results or cross-reactivity in a broad range of food matrixes containing no milk proteins. All participants obtained the BLG kit with a standard operational procedure, the list of the samples, samples, and a protocol for recording test results.The statistical tests (Cochran, Dixon, and Mandel) and analysis of variance were used to evaluate the interlaboratory study results. Repeatability and reproducibility limits, as well as LOQ (0.22 mg BLG/kg) and LOD (0.07 mg BLG/kg), for the kit were calculated.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
GM - Potravinářství
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/2B06138" target="_blank" >2B06138: Metody na stanovení alergenních složek v potravinách</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2009
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of AOAC International
ISSN
1060-3271
e-ISSN
—
Svazek periodika
92
Číslo periodika v rámci svazku
5
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
6
Strana od-do
—
Kód UT WoS článku
000271111200028