Standardisation of 56Co and 57Co using a software coincidence counting system
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F08%3A%230000288" target="_blank" >RIV/00177016:_____/08:#0000288 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Standardisation of 56Co and 57Co using a software coincidence counting system
Popis výsledku v původním jazyce
Activities of the radionuclides 56Co and 57Co were determined by the efficiency extrapolation method applied to 4pi(PC)-gamma coincidence counting. Solutions of 56Co usually contain a significant amount of 57Co and 58Co, so the measured activity of 56Corequires correction. When the conventional coincidence method is used for 56Co standardisation, the corrections are derived from the dependence of PC detection efficiencies for 57Co and 56Co measured using sources with mixture of 56Co and 57Co, which iscomplicated. These difficulties were reduced by means of a software coincidence method, with a HPGe detector in gamma channel, where the detection efficiencies were evaluated directly from the records of coincidence measurements of standardised sources.In the case of 57Co standardisation, the software coincidence counting system was applied for the evaluation of optimal setting of coincidence parameters.
Název v anglickém jazyce
Standardisation of 56Co and 57Co using a software coincidence counting system
Popis výsledku anglicky
Activities of the radionuclides 56Co and 57Co were determined by the efficiency extrapolation method applied to 4pi(PC)-gamma coincidence counting. Solutions of 56Co usually contain a significant amount of 57Co and 58Co, so the measured activity of 56Corequires correction. When the conventional coincidence method is used for 56Co standardisation, the corrections are derived from the dependence of PC detection efficiencies for 57Co and 56Co measured using sources with mixture of 56Co and 57Co, which iscomplicated. These difficulties were reduced by means of a software coincidence method, with a HPGe detector in gamma channel, where the detection efficiencies were evaluated directly from the records of coincidence measurements of standardised sources.In the case of 57Co standardisation, the software coincidence counting system was applied for the evaluation of optimal setting of coincidence parameters.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BG - Jaderná, atomová a molekulová fyzika, urychlovače
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
R - Projekt Ramcoveho programu EK
Ostatní
Rok uplatnění
2008
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Applied Radiation and Isotopes
ISSN
0969-8043
e-ISSN
—
Svazek periodika
66
Číslo periodika v rámci svazku
6-7
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
5
Strana od-do
—
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—