An automated method for monitoring aluminum in water samples based on a sequential injection platform
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11160%2F15%3A10301849" target="_blank" >RIV/00216208:11160/15:10301849 - isvavai.cz</a>
Výsledek na webu
<a href="http://pubs.rsc.org/en/content/articlehtml/2015/ay/c5ay01176k" target="_blank" >http://pubs.rsc.org/en/content/articlehtml/2015/ay/c5ay01176k</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1039/c5ay01176k" target="_blank" >10.1039/c5ay01176k</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
An automated method for monitoring aluminum in water samples based on a sequential injection platform
Popis výsledku v původním jazyce
The simple colour reaction between aluminum(III) and aluminon reagent was transferred to an automated sequential injection system. The experimental conditions were optimized in detail and found to be: sample pH, 4; zone sequence, aluminon-sample-HCl-aluminon; sample volume, 50 mu L; volume of 1 M HCl, 5 mu L; volume of reagent, 25 mu L. The interference caused by iron was eliminated by the addition of thioglycolic acid. The calibration plot was linear over the range of 100-800 mu g L-1 Al(III) with limits of detection and quantitation of 13.0 and 43.3 mu g L-1, respectively. The sample throughput was 36 samples per hour. The RSD values for repeatability tested on 200 and 600 mu g L-1 Al(III) levels were found in the range of 1.972.36% and 1.69-2.57%, respectively. Nine real samples of well and pond water were examined and the results corresponded well with the commercially available colour test based on the chromazurol reaction. The recovery based on the standard addition method ranged
Název v anglickém jazyce
An automated method for monitoring aluminum in water samples based on a sequential injection platform
Popis výsledku anglicky
The simple colour reaction between aluminum(III) and aluminon reagent was transferred to an automated sequential injection system. The experimental conditions were optimized in detail and found to be: sample pH, 4; zone sequence, aluminon-sample-HCl-aluminon; sample volume, 50 mu L; volume of 1 M HCl, 5 mu L; volume of reagent, 25 mu L. The interference caused by iron was eliminated by the addition of thioglycolic acid. The calibration plot was linear over the range of 100-800 mu g L-1 Al(III) with limits of detection and quantitation of 13.0 and 43.3 mu g L-1, respectively. The sample throughput was 36 samples per hour. The RSD values for repeatability tested on 200 and 600 mu g L-1 Al(III) levels were found in the range of 1.972.36% and 1.69-2.57%, respectively. Nine real samples of well and pond water were examined and the results corresponded well with the commercially available colour test based on the chromazurol reaction. The recovery based on the standard addition method ranged
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/EE2.3.20.0235" target="_blank" >EE2.3.20.0235: Vybudování výzkumného týmu experimentální a aplikované biofarmacie</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>S - Specificky vyzkum na vysokych skolach<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Analytical Methods
ISSN
1759-9660
e-ISSN
—
Svazek periodika
7
Číslo periodika v rámci svazku
13
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
8
Strana od-do
5530-5537
Kód UT WoS článku
000356959000024
EID výsledku v databázi Scopus
2-s2.0-84933544824