Determination of nitrophenols in drinking and river water by differential pulse voltammetry at boron-doped diamond film electrode.
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11310%2F11%3A10104944" target="_blank" >RIV/00216208:11310/11:10104944 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1002/elan.201100016" target="_blank" >http://dx.doi.org/10.1002/elan.201100016</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1002/elan.201100016" target="_blank" >10.1002/elan.201100016</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Determination of nitrophenols in drinking and river water by differential pulse voltammetry at boron-doped diamond film electrode.
Popis výsledku v původním jazyce
A differential pulse voltammetric method was developed for the determination of 2-nitrophenol, 4-nitrophenol and 2,4-dinitrophenol at a boron-doped diamond film electrode using electrochemical reduction and of 4-nitrophenol and 2,4-dinitrophenol using electrochemical oxidation. The method was successfully applied for the direct determination of these compounds in drinking and river water in the concentration range from 4 x 10(-7) to 2 x 10(-5) mol L(-1). To improve the limit of quantification, a preconcentration by solid phase extraction from 100 mL and 1000 mL of water samples was used with limit of determination around 2 x 10(-8) and 2 x 10(-7) mol L(-1), respectively.
Název v anglickém jazyce
Determination of nitrophenols in drinking and river water by differential pulse voltammetry at boron-doped diamond film electrode.
Popis výsledku anglicky
A differential pulse voltammetric method was developed for the determination of 2-nitrophenol, 4-nitrophenol and 2,4-dinitrophenol at a boron-doped diamond film electrode using electrochemical reduction and of 4-nitrophenol and 2,4-dinitrophenol using electrochemical oxidation. The method was successfully applied for the direct determination of these compounds in drinking and river water in the concentration range from 4 x 10(-7) to 2 x 10(-5) mol L(-1). To improve the limit of quantification, a preconcentration by solid phase extraction from 100 mL and 1000 mL of water samples was used with limit of determination around 2 x 10(-8) and 2 x 10(-7) mol L(-1), respectively.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CG - Elektrochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)<br>S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Electroanalysis
ISSN
1040-0397
e-ISSN
—
Svazek periodika
23
Číslo periodika v rámci svazku
5
Stát vydavatele periodika
DE - Spolková republika Německo
Počet stran výsledku
9
Strana od-do
1236-1244
Kód UT WoS článku
000290487600025