Separation efficiency of dual-selector systems in capillary electrophoresis
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11310%2F14%3A10189962" target="_blank" >RIV/00216208:11310/14:10189962 - isvavai.cz</a>
Výsledek na webu
<a href="http://ac.els-cdn.com/S0021967314000296/1-s2.0-S0021967314000296-main.pdf?_tid=e994d2bc-a561-11e4-80e7-00000aab0f6b&acdnat=1422280247_11ddaa9da1d9edef54a04e4866c0c8da" target="_blank" >http://ac.els-cdn.com/S0021967314000296/1-s2.0-S0021967314000296-main.pdf?_tid=e994d2bc-a561-11e4-80e7-00000aab0f6b&acdnat=1422280247_11ddaa9da1d9edef54a04e4866c0c8da</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.chroma.2014.01.006" target="_blank" >10.1016/j.chroma.2014.01.006</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Separation efficiency of dual-selector systems in capillary electrophoresis
Popis výsledku v původním jazyce
We introduce an easy but highly descriptive model of separation efficiency of dual-selector systems in capillary electrophoresis.The model expresses effective mobilities of analytes in dual-selector mix-tures as a function of mixture composition and total concentration. The effective mobility follows the pattern familiar from single-selector systems, while complexation constant and mobility of the complex are replaced by the same but "overall" parameters and a total concentration of the mixture takes the role of a selector concentration. The overall parameters can be either calculated from the individual ones (an arbitrary mixture) or measured directly (a particular mixture). We inspected two model dual-selector systems consisting of heptakis(2,6-di-O-methyl)- -CD and -CD and of heptakis(2,6-di- O-methyl)- -CD and 6-O- -maltosyl- -CD, and ibuprofen and flurbiprofen as model analytes (pH 8.2, non-enantioselective separation).Adopting any optimization strategy typically used in single-se
Název v anglickém jazyce
Separation efficiency of dual-selector systems in capillary electrophoresis
Popis výsledku anglicky
We introduce an easy but highly descriptive model of separation efficiency of dual-selector systems in capillary electrophoresis.The model expresses effective mobilities of analytes in dual-selector mix-tures as a function of mixture composition and total concentration. The effective mobility follows the pattern familiar from single-selector systems, while complexation constant and mobility of the complex are replaced by the same but "overall" parameters and a total concentration of the mixture takes the role of a selector concentration. The overall parameters can be either calculated from the individual ones (an arbitrary mixture) or measured directly (a particular mixture). We inspected two model dual-selector systems consisting of heptakis(2,6-di-O-methyl)- -CD and -CD and of heptakis(2,6-di- O-methyl)- -CD and 6-O- -maltosyl- -CD, and ibuprofen and flurbiprofen as model analytes (pH 8.2, non-enantioselective separation).Adopting any optimization strategy typically used in single-se
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GPP206%2F12%2FP630" target="_blank" >GPP206/12/P630: Sjednocený model (dynamické) chirální separace v kapilární zónové elektroforéze a integrované softwarové řešení</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)<br>S - Specificky vyzkum na vysokych skolach<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Chromatography A
ISSN
0021-9673
e-ISSN
—
Svazek periodika
1330
Číslo periodika v rámci svazku
February 2014
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
7
Strana od-do
82-88