Elektrochemický senzor pro detekci acetylsalicylové kyseliny
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14310%2F06%3A00016396" target="_blank" >RIV/00216224:14310/06:00016396 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/62156489:43210/06:00103425 RIV/62157124:16370/06:00000751 RIV/62157124:16270/06:00000867
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Electrochemical sensors for detection of acetylsalicylic acid
Popis výsledku v původním jazyce
Acetylsalicylic acid (AcSA), or aspirin, was introduced in the late 1890s and has been used to treat a variety of inflammatory conditions. The aim of this work was to suggest electrochemical sensor for acetylsalicylic detection. Primarily, we utilized square wave voltammetry (SWV) using both carbon paste electrode (CPE) and of graphite pencil electrode (GPE) as working ones to indirect determination of AcSA. The principle of indirect determination of AcSA bases in its hydrolysis on salicylic acid (SA),which is consequently detected. Thus, we optimized both determination of SA and conditions for AcSA hydrolysis and found out that the most suitable frequency, amplitude, step potential and the composition and pH of the supporting electrolyte for the determination of SA was 260 Hz, 50 mV, 10 mV and Britton-Robinson buffer (pH 1.81), respectively. The detection limit (S/N = 3) of the SA was 1.3 ng/ml. After that, we aimed on indirect determination of AcSA by SWV CPE. We tested the influenc
Název v anglickém jazyce
Electrochemical sensors for detection of acetylsalicylic acid
Popis výsledku anglicky
Acetylsalicylic acid (AcSA), or aspirin, was introduced in the late 1890s and has been used to treat a variety of inflammatory conditions. The aim of this work was to suggest electrochemical sensor for acetylsalicylic detection. Primarily, we utilized square wave voltammetry (SWV) using both carbon paste electrode (CPE) and of graphite pencil electrode (GPE) as working ones to indirect determination of AcSA. The principle of indirect determination of AcSA bases in its hydrolysis on salicylic acid (SA),which is consequently detected. Thus, we optimized both determination of SA and conditions for AcSA hydrolysis and found out that the most suitable frequency, amplitude, step potential and the composition and pH of the supporting electrolyte for the determination of SA was 260 Hz, 50 mV, 10 mV and Britton-Robinson buffer (pH 1.81), respectively. The detection limit (S/N = 3) of the SA was 1.3 ng/ml. After that, we aimed on indirect determination of AcSA by SWV CPE. We tested the influenc
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CG - Elektrochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2006
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
SENSORS
ISSN
1424-8220
e-ISSN
—
Svazek periodika
11
Číslo periodika v rámci svazku
6
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
15
Strana od-do
1483-1497
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—