LA-ICP-MS analysis of metal layers on samples of cultural heritage
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14310%2F19%3A00110379" target="_blank" >RIV/00216224:14310/19:00110379 - isvavai.cz</a>
Výsledek na webu
<a href="https://link.springer.com/article/10.1007/s11696-019-00745-6" target="_blank" >https://link.springer.com/article/10.1007/s11696-019-00745-6</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1007/s11696-019-00745-6" target="_blank" >10.1007/s11696-019-00745-6</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
LA-ICP-MS analysis of metal layers on samples of cultural heritage
Popis výsledku v původním jazyce
LA-ICP-MS (laser ablation inductively coupled plasma mass spectrometry) was applied as a semi-destructive method for the analysis of metal layers on samples of cultural heritage. The LA system operates at a wavelength of 193 nm with a pulse duration of 4 ns. The laser ablation was performed using spot mode with a diameter of 110 mu m. This method was developed using eight varied reference samples with a defined layer depth and composition. Four real samples of watch cases were analysed, and the layer composition and thickness were determined together with two test reference samples. The analytical results were compared with X-ray fluorescence spectroscopy analysis. The layer thickness was additionally determined by scanning electron microscopy on the cross section of selected samples.
Název v anglickém jazyce
LA-ICP-MS analysis of metal layers on samples of cultural heritage
Popis výsledku anglicky
LA-ICP-MS (laser ablation inductively coupled plasma mass spectrometry) was applied as a semi-destructive method for the analysis of metal layers on samples of cultural heritage. The LA system operates at a wavelength of 193 nm with a pulse duration of 4 ns. The laser ablation was performed using spot mode with a diameter of 110 mu m. This method was developed using eight varied reference samples with a defined layer depth and composition. Four real samples of watch cases were analysed, and the layer composition and thickness were determined together with two test reference samples. The analytical results were compared with X-ray fluorescence spectroscopy analysis. The layer thickness was additionally determined by scanning electron microscopy on the cross section of selected samples.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10400 - Chemical sciences
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2019
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Chemical Papers
ISSN
2585-7290
e-ISSN
1336-9075
Svazek periodika
73
Číslo periodika v rámci svazku
12
Stát vydavatele periodika
CH - Švýcarská konfederace
Počet stran výsledku
14
Strana od-do
2923-2936
Kód UT WoS článku
000488930200006
EID výsledku v databázi Scopus
2-s2.0-85073209463