Antimony Trifluoride-Modified Carbon Paste Electrodes for Electrochemical Stripping Analysis of Selected Heavy Metals
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216275%3A25310%2F11%3A39892683" target="_blank" >RIV/00216275:25310/11:39892683 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Antimony Trifluoride-Modified Carbon Paste Electrodes for Electrochemical Stripping Analysis of Selected Heavy Metals
Popis výsledku v původním jazyce
In this article, a new typ of non-mercury metal-based electrode, antimony trifluoride-bulk-modified carbon paste electrode (SbF3-CPE) is for the first time reported and examined for electrochemical stripping analysis of selected heavy metal ions at theirtrace concentration level. In the role of bulk modifier and a source of antimony film generated in state nascenti, SbF3 in a content of 3 % (w/w) in the carbon paste mixture was the ultimate choice. All the important experimental parameters have been optimized by using Cd(II) and Pb(II) as the model ions and employing the square-wave anodic stripping voltammetry (SWASV) as the proper technique. As found out, the SbF3-CPE has exhibited a finely linear response (with R2 = 0.999) in the concentration range of 20-120 mýg L-1 Me(II), while the signal of interest could be reproduced with a relative error of 6.5 % for Cd and 1.2 % for Pb (when n = 7).
Název v anglickém jazyce
Antimony Trifluoride-Modified Carbon Paste Electrodes for Electrochemical Stripping Analysis of Selected Heavy Metals
Popis výsledku anglicky
In this article, a new typ of non-mercury metal-based electrode, antimony trifluoride-bulk-modified carbon paste electrode (SbF3-CPE) is for the first time reported and examined for electrochemical stripping analysis of selected heavy metal ions at theirtrace concentration level. In the role of bulk modifier and a source of antimony film generated in state nascenti, SbF3 in a content of 3 % (w/w) in the carbon paste mixture was the ultimate choice. All the important experimental parameters have been optimized by using Cd(II) and Pb(II) as the model ions and employing the square-wave anodic stripping voltammetry (SWASV) as the proper technique. As found out, the SbF3-CPE has exhibited a finely linear response (with R2 = 0.999) in the concentration range of 20-120 mýg L-1 Me(II), while the signal of interest could be reproduced with a relative error of 6.5 % for Cd and 1.2 % for Pb (when n = 7).
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
CG - Elektrochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/LC06035" target="_blank" >LC06035: Centrum biofyzikální chemie, bioelektrochemie a bioanalýzy. Nové nástroje pro genomiku, proteomiku a biomedicínu.</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Sensing in Electroanalysis Vol. 6
ISBN
978-80-7395-434-5
ISSN
—
e-ISSN
—
Počet stran výsledku
14
Strana od-do
205-218
Název nakladatele
Univerzita Pardubice
Místo vydání
Pardubice
Místo konání akce
Pardubice
Datum konání akce
14. 11. 2011
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—