Determination of Phosphates and Polyphosphates Using Capillary Isotachophoresis
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216275%3A25310%2F12%3A39895542" target="_blank" >RIV/00216275:25310/12:39895542 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Determination of Phosphates and Polyphosphates Using Capillary Isotachophoresis
Popis výsledku v původním jazyce
In this article, phosphate, di- and triphosphates are the analytes determined in commercially marketed samples with the aid of capillary isotachophoresis. After method development, the optimal electrolyte system comprised 10 mM HCl + beta-alanine + 0.1%hydroxy-ethylcellulose (pH 3.6) as the leading electrolyte and 10 mM citric acid as the terminating electrolyte. The calibration / detection characteristics are as follows: linearity over the concentration range of 0-200 mg.l-1 P2O5, limit of detection ca. 0.3 mg.l-1 P2O5, limit of quantification of ca. 1 mg.l-1 P2O5. The time of analysis was from 15 to 30 minutes. Furthermore, relatively simple procedure, low operational costs, and sufficient sensitivity are further positive attributes of the method tested in analyses of selected real samples.
Název v anglickém jazyce
Determination of Phosphates and Polyphosphates Using Capillary Isotachophoresis
Popis výsledku anglicky
In this article, phosphate, di- and triphosphates are the analytes determined in commercially marketed samples with the aid of capillary isotachophoresis. After method development, the optimal electrolyte system comprised 10 mM HCl + beta-alanine + 0.1%hydroxy-ethylcellulose (pH 3.6) as the leading electrolyte and 10 mM citric acid as the terminating electrolyte. The calibration / detection characteristics are as follows: linearity over the concentration range of 0-200 mg.l-1 P2O5, limit of detection ca. 0.3 mg.l-1 P2O5, limit of quantification of ca. 1 mg.l-1 P2O5. The time of analysis was from 15 to 30 minutes. Furthermore, relatively simple procedure, low operational costs, and sufficient sensitivity are further positive attributes of the method tested in analyses of selected real samples.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/EE2.3.30.0021" target="_blank" >EE2.3.30.0021: Posílení excelentních týmů výzkumu a vývoje na Univerzitě Pardubice</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Sensing in Electroanalysis Vol. 7
ISBN
978-80-7395-563-2
ISSN
—
e-ISSN
—
Počet stran výsledku
10
Strana od-do
423-432
Název nakladatele
Univerzita Pardubice
Místo vydání
Pardubice
Místo konání akce
Pardubice
Datum konání akce
14. 11. 2012
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—