Determination of Methotrexate at a Silver Solid Amalgam Electrode by Differential Pulse Voltammetry
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216275%3A25310%2F16%3A39901439" target="_blank" >RIV/00216275:25310/16:39901439 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1080/00032719.2014.996812" target="_blank" >http://dx.doi.org/10.1080/00032719.2014.996812</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1080/00032719.2014.996812" target="_blank" >10.1080/00032719.2014.996812</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Determination of Methotrexate at a Silver Solid Amalgam Electrode by Differential Pulse Voltammetry
Popis výsledku v původním jazyce
An ultrasensitive voltammetric method is reported for the determination of methotrexate using a polished modified silver solid amalgam electrode. Electrochemical behavior of this compound was examined using cyclic, linear sweep, and differential pulse voltammetry. Methotrexate provided two reduction and one oxidation signal in acidic and neutral media. The first reduction peak (E-p=-350 millivolts vs. Ag|AgCl electrode) and supporting electrolyte of pH 5 was chosen for further experiments. Differential pulse voltammetry parameters were optimized in combination with adsorptive stripping and the applicability of the method was verified. A low limit of detection (1.5x10(-10)moles per liter), good repeatability of determination (relative standard deviation<3.5%, n=5), and good repeatability of measurements (relative standard deviation of 1.7% at 1.0x10(-7)mole per liter methotrexate and n=11) was obtained. The method was employed for the accurate determination of methotrexate in a chemotherapeutic preparation.
Název v anglickém jazyce
Determination of Methotrexate at a Silver Solid Amalgam Electrode by Differential Pulse Voltammetry
Popis výsledku anglicky
An ultrasensitive voltammetric method is reported for the determination of methotrexate using a polished modified silver solid amalgam electrode. Electrochemical behavior of this compound was examined using cyclic, linear sweep, and differential pulse voltammetry. Methotrexate provided two reduction and one oxidation signal in acidic and neutral media. The first reduction peak (E-p=-350 millivolts vs. Ag|AgCl electrode) and supporting electrolyte of pH 5 was chosen for further experiments. Differential pulse voltammetry parameters were optimized in combination with adsorptive stripping and the applicability of the method was verified. A low limit of detection (1.5x10(-10)moles per liter), good repeatability of determination (relative standard deviation<3.5%, n=5), and good repeatability of measurements (relative standard deviation of 1.7% at 1.0x10(-7)mole per liter methotrexate and n=11) was obtained. The method was employed for the accurate determination of methotrexate in a chemotherapeutic preparation.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2016
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Analytical Letters
ISSN
0003-2719
e-ISSN
—
Svazek periodika
49
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
13
Strana od-do
122-134
Kód UT WoS článku
000363048900011
EID výsledku v databázi Scopus
—