Vysoká rozlišovací schopnost ultrazvukových obrazů pomocí signálu 1. a 2. harmonické
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F04%3APU37534" target="_blank" >RIV/00216305:26220/04:PU37534 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Superresolution of Ultrasound Images Using the 1st and 2nd Harmonic Signal
Popis výsledku v původním jazyce
This paper presents a new method of blind 2-D homomorphic deconvolution and speckle reduction applied to medical ultrasound images. The deconvolution technique is based on an improved 2-D phase unwrapping scheme for pulse estimation. The input images aredecomposed into minimum-phase and allpass components. The 2-D phase unwrapping is applied only to the allpass component. The 2-D phase of the minimum-phase component is derived by a Hilbert transform. The accuracy of 2-D phase unnwrapping is also improved by processing small (16x16 pixels) overlapping subimages separately. This takes the spatial variance of the ultrasound pulse into account. The deconvolution algorithm is applied separately to the 1st and 2nd harmonic images, producing much sharper images of approximately the same resolution and different speckle patterns. Speckle reduction is made by adding the envelope images of the deconvolved 1st and 2nd harmonic images. It does neither decrease the
Název v anglickém jazyce
Superresolution of Ultrasound Images Using the 1st and 2nd Harmonic Signal
Popis výsledku anglicky
This paper presents a new method of blind 2-D homomorphic deconvolution and speckle reduction applied to medical ultrasound images. The deconvolution technique is based on an improved 2-D phase unwrapping scheme for pulse estimation. The input images aredecomposed into minimum-phase and allpass components. The 2-D phase unwrapping is applied only to the allpass component. The 2-D phase of the minimum-phase component is derived by a Hilbert transform. The accuracy of 2-D phase unnwrapping is also improved by processing small (16x16 pixels) overlapping subimages separately. This takes the spatial variance of the ultrasound pulse into account. The deconvolution algorithm is applied separately to the 1st and 2nd harmonic images, producing much sharper images of approximately the same resolution and different speckle patterns. Speckle reduction is made by adding the envelope images of the deconvolved 1st and 2nd harmonic images. It does neither decrease the
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JD - Využití počítačů, robotika a její aplikace
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA102%2F02%2F0890" target="_blank" >GA102/02/0890: Analýza medicínských ultrasonografických dat se zaměřením na 3D zobrazení v kardiologii</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2004
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
IEEE Transactions on Ultrasonocs, Ferroelectrics, and Frequency Control
ISSN
0885-3010
e-ISSN
—
Svazek periodika
51
Číslo periodika v rámci svazku
2
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
13
Strana od-do
163-175
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—