Noise Contact Study of CdTe Radiation Detectors
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F12%3APU101449" target="_blank" >RIV/00216305:26220/12:PU101449 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Noise Contact Study of CdTe Radiation Detectors
Popis výsledku v původním jazyce
We have compared noise spectra and charge transport properties analysis results of CdTe radiation detectors: low-ohmic and semi-insulating. We observed assymmetry of IV characteristics of the low-ohmic detector between "normal" nad "reverse" bias, showing improper quality of contacts preparation, which resulted in higher concentration of impurities in M-S interface area. This finding was supported by the fact that the low frequency noise spectral density magnitude was proportional to applied voltage with exponent of 2.7, which is higher than theoretical value 2. The more advanced manufacturing technology of the semi-insulating detector resulted in symmetric contacts IV characteristics and its linear character. Also, this detector showed lower increaseof noise spectral density magnitude with exponent of 2.3.
Název v anglickém jazyce
Noise Contact Study of CdTe Radiation Detectors
Popis výsledku anglicky
We have compared noise spectra and charge transport properties analysis results of CdTe radiation detectors: low-ohmic and semi-insulating. We observed assymmetry of IV characteristics of the low-ohmic detector between "normal" nad "reverse" bias, showing improper quality of contacts preparation, which resulted in higher concentration of impurities in M-S interface area. This finding was supported by the fact that the low frequency noise spectral density magnitude was proportional to applied voltage with exponent of 2.7, which is higher than theoretical value 2. The more advanced manufacturing technology of the semi-insulating detector resulted in symmetric contacts IV characteristics and its linear character. Also, this detector showed lower increaseof noise spectral density magnitude with exponent of 2.3.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
BM - Fyzika pevných látek a magnetismus
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of the Scientific Conference Physics of Materials 2012
ISBN
978-80-553-1175-3
ISSN
—
e-ISSN
—
Počet stran výsledku
6
Strana od-do
99-104
Název nakladatele
Neuveden
Místo vydání
Košice, Slovensko
Místo konání akce
Košice
Datum konání akce
17. 10. 2012
Typ akce podle státní příslušnosti
WRD - Celosvětová akce