Signal processing of 3D reaction field in vicinity of non-ferromagnetic specimen and its magnetic susceptibility calculation
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F12%3APU97553" target="_blank" >RIV/00216305:26220/12:PU97553 - isvavai.cz</a>
Výsledek na webu
<a href="http://elektrorevue.cz/en/articles/analogue-technics/0/signal-processing-of-3d-reaction-field-in-vicinity-of-non-ferromagnetic-specimen-and-its-magnetic-susceptibility-calculation/" target="_blank" >http://elektrorevue.cz/en/articles/analogue-technics/0/signal-processing-of-3d-reaction-field-in-vicinity-of-non-ferromagnetic-specimen-and-its-magnetic-susceptibility-calculation/</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Signal processing of 3D reaction field in vicinity of non-ferromagnetic specimen and its magnetic susceptibility calculation
Popis výsledku v původním jazyce
To measure magnetic susceptibility several methods are employed that use regular-shape specimens of diamagnetic and paramagnetic substances. This paper describes an MRI method for measuring the magnetic susceptibility of specimens in the shape of cylindrical bars of arbitrary diameter that do not produce an MR signal. The proposed method is based on measuring a 3D map of the reaction field in the vicinity of the specimen. This reaction field was mapped by Gradient Echo method (GE). The method can be used for specimens of irregular cross-section and it is not necessary to use mathematical simulation of magnetic field distribution. A great advantage of our method for susceptibility evaluation consists in using only standard commercially used devices without needing any special sequence.
Název v anglickém jazyce
Signal processing of 3D reaction field in vicinity of non-ferromagnetic specimen and its magnetic susceptibility calculation
Popis výsledku anglicky
To measure magnetic susceptibility several methods are employed that use regular-shape specimens of diamagnetic and paramagnetic substances. This paper describes an MRI method for measuring the magnetic susceptibility of specimens in the shape of cylindrical bars of arbitrary diameter that do not produce an MR signal. The proposed method is based on measuring a 3D map of the reaction field in the vicinity of the specimen. This reaction field was mapped by Gradient Echo method (GE). The method can be used for specimens of irregular cross-section and it is not necessary to use mathematical simulation of magnetic field distribution. A great advantage of our method for susceptibility evaluation consists in using only standard commercially used devices without needing any special sequence.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Elektrorevue - Internetový časopis (http://www.elektrorevue.cz)
ISSN
1213-1539
e-ISSN
—
Svazek periodika
2012 (3)
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
6
Strana od-do
14-19
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—