DESIGN OF THE PULSE GENERATOR FOR TESTING AND VALIDATING PARTIAL DISCHARGE MEASURING SYSTEMS
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F21%3APU141480" target="_blank" >RIV/00216305:26220/21:PU141480 - isvavai.cz</a>
Výsledek na webu
<a href="https://www.fekt.vut.cz/conf/EEICT/archiv/sborniky/EEICT_2021_sbornik_1.pdf" target="_blank" >https://www.fekt.vut.cz/conf/EEICT/archiv/sborniky/EEICT_2021_sbornik_1.pdf</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
DESIGN OF THE PULSE GENERATOR FOR TESTING AND VALIDATING PARTIAL DISCHARGE MEASURING SYSTEMS
Popis výsledku v původním jazyce
This article deals with the design and implementation of a pulse generator serving as a cali-bration generator for setting up measuring systems for monitoring the activity of partial dis-charges (PD) in the high-voltage power transformers or gas insulated switchgears (GIS). After installing the measuring device, it is advisable to test the entire measuring chain and, if neces-sary, set parameters that would affect the correct detection of PD signals. Calibration pulse generators are used for setting and validating measuring chain. These are usually used in sys-tems that sense the electromagnetic activity of PD. Correct and timely detection of PD can avert a transformer accident and thus contribute to higher safety in power plants. The pro-posed generator should serve as the output part of a programmable generator, which did not reach sufficient parameters of output pulses. The goal was to achieve a rising time around 100 ps and an amplitude of more than 6 V using modern and very fast operational ampli
Název v anglickém jazyce
DESIGN OF THE PULSE GENERATOR FOR TESTING AND VALIDATING PARTIAL DISCHARGE MEASURING SYSTEMS
Popis výsledku anglicky
This article deals with the design and implementation of a pulse generator serving as a cali-bration generator for setting up measuring systems for monitoring the activity of partial dis-charges (PD) in the high-voltage power transformers or gas insulated switchgears (GIS). After installing the measuring device, it is advisable to test the entire measuring chain and, if neces-sary, set parameters that would affect the correct detection of PD signals. Calibration pulse generators are used for setting and validating measuring chain. These are usually used in sys-tems that sense the electromagnetic activity of PD. Correct and timely detection of PD can avert a transformer accident and thus contribute to higher safety in power plants. The pro-posed generator should serve as the output part of a programmable generator, which did not reach sufficient parameters of output pulses. The goal was to achieve a rising time around 100 ps and an amplitude of more than 6 V using modern and very fast operational ampli
Klasifikace
Druh
O - Ostatní výsledky
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2021
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů