High pressure modification of the simple automated liquid chromatographic system for splitless nano column gradient separations
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26310%2F13%3APU106482" target="_blank" >RIV/00216305:26310/13:PU106482 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
High pressure modification of the simple automated liquid chromatographic system for splitless nano column gradient separations
Popis výsledku v původním jazyce
In this work we have modified our concept of simple splitless gradient liquid chromatographic system for micro and nano columns with backpressure up to 50 bar. Presented configuration allows us to run analysis under the pressure up to 400 bar. Experimental configuration consists of an OEM programmable syringe pump equipped with a 250-ul high-pressure syringe and ten-port selector valve. High pressure syringe is connected to central port of selector valve via 0.5 meter long PEEK capillary of 380 um i. d.fixed in horizontal plane. PEEK capillary here serves as gradient generator as well as gradient loop and as sample loop. Mobile phase gradients are created by successive sucking of defined volumes of four water-acetonitrile solutions with gradually changing acetonitrile concentration into the PEEK capillary. Sharp mobile phase boundaries fade out while passing through the valve body and due to the parabolic profile of laminar flow as well as due to diffusion. Available linear gradients
Název v anglickém jazyce
High pressure modification of the simple automated liquid chromatographic system for splitless nano column gradient separations
Popis výsledku anglicky
In this work we have modified our concept of simple splitless gradient liquid chromatographic system for micro and nano columns with backpressure up to 50 bar. Presented configuration allows us to run analysis under the pressure up to 400 bar. Experimental configuration consists of an OEM programmable syringe pump equipped with a 250-ul high-pressure syringe and ten-port selector valve. High pressure syringe is connected to central port of selector valve via 0.5 meter long PEEK capillary of 380 um i. d.fixed in horizontal plane. PEEK capillary here serves as gradient generator as well as gradient loop and as sample loop. Mobile phase gradients are created by successive sucking of defined volumes of four water-acetonitrile solutions with gradually changing acetonitrile concentration into the PEEK capillary. Sharp mobile phase boundaries fade out while passing through the valve body and due to the parabolic profile of laminar flow as well as due to diffusion. Available linear gradients
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2013
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Chemické listy
ISSN
1213-7103
e-ISSN
—
Svazek periodika
107
Číslo periodika v rámci svazku
s3
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
3
Strana od-do
438-440