Amplitude envelope slopes of oscillometric blood pressure waveforms as defined by amplitude ratios
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F49777513%3A23220%2F09%3A00502251" target="_blank" >RIV/49777513:23220/09:00502251 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Amplitude envelope slopes of oscillometric blood pressure waveforms as defined by amplitude ratios
Popis výsledku v původním jazyce
Oscillometric methods for blood pressure determination use oscillometric waveform amplitudes. A typical waveform amplitude envelope exhibits a rise of amplitudes at cuff pressures above mean arterial pressure and decrease of amplitudes below mean pressure. Amplitude envelope slopes are functions of several hemodynamic variables. An amplitude envelope can be divided into three segments: segment S1 is the portion of the slope at cuff pressures above systolic pressure, segment S2 is the portion at cuff pressures between systolic and mean pressure, and segment S3 is the slope between mean and diastolic pressure. The slope S2 should be less steep than slope S1 or slope S3 because of the blood flow under the cuff and into the distal vasculature. To test thishypothesis, ten subjects were tested using a specialized data acquisition system. Commercial brachial (upper arm) cuff was used. Results (mean values): S1=0.125, S2=0.056, S3=0.093. All ten S2 slopes were less steep than the slopes S1. T
Název v anglickém jazyce
Amplitude envelope slopes of oscillometric blood pressure waveforms as defined by amplitude ratios
Popis výsledku anglicky
Oscillometric methods for blood pressure determination use oscillometric waveform amplitudes. A typical waveform amplitude envelope exhibits a rise of amplitudes at cuff pressures above mean arterial pressure and decrease of amplitudes below mean pressure. Amplitude envelope slopes are functions of several hemodynamic variables. An amplitude envelope can be divided into three segments: segment S1 is the portion of the slope at cuff pressures above systolic pressure, segment S2 is the portion at cuff pressures between systolic and mean pressure, and segment S3 is the slope between mean and diastolic pressure. The slope S2 should be less steep than slope S1 or slope S3 because of the blood flow under the cuff and into the distal vasculature. To test thishypothesis, ten subjects were tested using a specialized data acquisition system. Commercial brachial (upper arm) cuff was used. Results (mean values): S1=0.125, S2=0.056, S3=0.093. All ten S2 slopes were less steep than the slopes S1. T
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/FT-TA3%2F058" target="_blank" >FT-TA3/058: *Signálový procesor pro zpracování radarových dat.</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2009
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
2009 Applied Electronics
ISBN
978-80-7043-781-0
ISSN
—
e-ISSN
—
Počet stran výsledku
4
Strana od-do
—
Název nakladatele
University of West Bohemia
Místo vydání
Pilsen
Místo konání akce
Plzeň
Datum konání akce
10. 9. 2009
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000273977300029