Kombinace plynové chromatografie s hmotnostní spektrometrií v proudové trubici s vybranými ionty pro absolutní kvantifikaci stopových plynů
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61388955%3A_____%2F06%3A00078424" target="_blank" >RIV/61388955:_____/06:00078424 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Combined use of gas chromatography and selected ion flow tube mass spectrometry for absolute trace gas quantification
Popis výsledku v původním jazyce
The value of the gas chromatography (GC) and selected ion flow tube mass spectrometry (SIFT-MS) combination for the analysis of trace gases is demonstrated by the quantification of acetone in air samples using the three precursor ions available to SIFT-MS, viz. H3O+, NO+ and O-2(+), and by the separation of the isomers 1-propanol and 2-propanol, and their analysis using H3O+ precursor ions. It is shown that the GC/SIFT-MS combination allows for accurate trace gas quantification obviating the regular, time-consuming calibrations that are usually required for the more commonly used detectors of GC systems, and the positive identification of isomers in mixtures that is often challenging using SIFT-MS alone. Thus, the GC/SIFT-MS combination paves the way to more confident analyses of complex mixtures such as exhaled breath.
Název v anglickém jazyce
Combined use of gas chromatography and selected ion flow tube mass spectrometry for absolute trace gas quantification
Popis výsledku anglicky
The value of the gas chromatography (GC) and selected ion flow tube mass spectrometry (SIFT-MS) combination for the analysis of trace gases is demonstrated by the quantification of acetone in air samples using the three precursor ions available to SIFT-MS, viz. H3O+, NO+ and O-2(+), and by the separation of the isomers 1-propanol and 2-propanol, and their analysis using H3O+ precursor ions. It is shown that the GC/SIFT-MS combination allows for accurate trace gas quantification obviating the regular, time-consuming calibrations that are usually required for the more commonly used detectors of GC systems, and the positive identification of isomers in mixtures that is often challenging using SIFT-MS alone. Thus, the GC/SIFT-MS combination paves the way to more confident analyses of complex mixtures such as exhaled breath.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CF - Fyzikální chemie a teoretická chemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA202%2F03%2F0827" target="_blank" >GA202/03/0827: Studium elementárních procesů v nízkoteplotním a technologicky orientovaném plazmatu a vývoj relevantních diagnostických metod</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2006
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Rapid Communications in Mass Spectrometry
ISSN
0951-4198
e-ISSN
—
Svazek periodika
20
Číslo periodika v rámci svazku
4
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
5
Strana od-do
563-567
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—