Potenciometrický detektor heparinu: přechodové chování a mechanismus odezvy
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61388955%3A_____%2F07%3A00081281" target="_blank" >RIV/61388955:_____/07:00081281 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Potentiometric Sensor for Heparin Polyion: Transient Behavior and Response Mechanism
Popis výsledku v původním jazyce
Chronopotentiometry and electrochemical impedance spectroscopy were used to study the transient behavior and the potentiometric response mechanism of the polymer membrane-based sensor for heparin. Membrane with a composition of 66 wt % poly(vinyl chloride), 33 wt % o-nitrophenyl octyl ether (plasticizer), and 0.05 M tridodecylmethylammonium chloride (ion exchanger) was deposited on the surface of a silver or a glassy carbon (GC) electrode. In the latter case, the membrane contained also 0.1 M 1,1'-dimethylferrocene/1,1'-dimethylferricenium(+) couple ensuring the electronic contact between the membrane and GC. The sensor was dipped in an aqueous solution of 0.1 M LiCl, which was stirred with a magnetic stirrer (2-18.2 Hz), and eventually spiked with heparin (0.05-5 U mL(-1)). Chronopotentiometric measurements were carried out using either the Ag supported membrane with a thickness > 100 mu m or the GC supported membrane with a defined thickness of 2-30 mu m, ...
Název v anglickém jazyce
Potentiometric Sensor for Heparin Polyion: Transient Behavior and Response Mechanism
Popis výsledku anglicky
Chronopotentiometry and electrochemical impedance spectroscopy were used to study the transient behavior and the potentiometric response mechanism of the polymer membrane-based sensor for heparin. Membrane with a composition of 66 wt % poly(vinyl chloride), 33 wt % o-nitrophenyl octyl ether (plasticizer), and 0.05 M tridodecylmethylammonium chloride (ion exchanger) was deposited on the surface of a silver or a glassy carbon (GC) electrode. In the latter case, the membrane contained also 0.1 M 1,1'-dimethylferrocene/1,1'-dimethylferricenium(+) couple ensuring the electronic contact between the membrane and GC. The sensor was dipped in an aqueous solution of 0.1 M LiCl, which was stirred with a magnetic stirrer (2-18.2 Hz), and eventually spiked with heparin (0.05-5 U mL(-1)). Chronopotentiometric measurements were carried out using either the Ag supported membrane with a thickness > 100 mu m or the GC supported membrane with a defined thickness of 2-30 mu m, ...
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CG - Elektrochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA203%2F04%2F0424" target="_blank" >GA203/04/0424: Elektrochemie biomedicinálně významných polyiontů na rozhraní dvou nemísitelných roztoků elektrolytů s aplikacemi v analýze léčiv</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2007
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Analytical Chemistry
ISSN
0793-0135
e-ISSN
—
Svazek periodika
79
Číslo periodika v rámci svazku
7
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
9
Strana od-do
2892-2900
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—