On the features, successes and challenges of selected ion flow tube mass spectrometry
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61388955%3A_____%2F13%3A00421167" target="_blank" >RIV/61388955:_____/13:00421167 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1255/ejms.1240" target="_blank" >http://dx.doi.org/10.1255/ejms.1240</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1255/ejms.1240" target="_blank" >10.1255/ejms.1240</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
On the features, successes and challenges of selected ion flow tube mass spectrometry
Popis výsledku v původním jazyce
The major features of the selected ion flow tube mass spectrometry (SIFT-MS) analytical method that was conceived and designed for the analysis, in real time, of air obviating sample collections into bags or extraction by pre-concentration of trace compounds onto surfaces are reviewed. The unique analytical capabilities of SIFT-MS for ambient analysis are stressed that allow quantification of volatile organic and inorganic compounds directly from the measurement of physical parameters without the need for regular instrumental calibration using internal or external standards. Then, emphasis is placed on the challenging real-time accurate analysis of single exhalations of humid breath, which is now achieved and readily facilitates wider applications of SIFT-MS in other fields where trace gas analysis has value. The quality of the data obtained by SIFT-MS is illustrated by the quantification of some exhaled breath metabolites that are of immediate relevance to physiology and medicine, inc
Název v anglickém jazyce
On the features, successes and challenges of selected ion flow tube mass spectrometry
Popis výsledku anglicky
The major features of the selected ion flow tube mass spectrometry (SIFT-MS) analytical method that was conceived and designed for the analysis, in real time, of air obviating sample collections into bags or extraction by pre-concentration of trace compounds onto surfaces are reviewed. The unique analytical capabilities of SIFT-MS for ambient analysis are stressed that allow quantification of volatile organic and inorganic compounds directly from the measurement of physical parameters without the need for regular instrumental calibration using internal or external standards. Then, emphasis is placed on the challenging real-time accurate analysis of single exhalations of humid breath, which is now achieved and readily facilitates wider applications of SIFT-MS in other fields where trace gas analysis has value. The quality of the data obtained by SIFT-MS is illustrated by the quantification of some exhaled breath metabolites that are of immediate relevance to physiology and medicine, inc
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CF - Fyzikální chemie a teoretická chemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA13-28882S" target="_blank" >GA13-28882S: Excitace, separace a fragmentace iontů v driftovém poli v SIFT-MS pro okamžitou selektivní kvantifikaci stopových plynů</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2013
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
European Journal of Mass Spectrometry
ISSN
1469-0667
e-ISSN
—
Svazek periodika
19
Číslo periodika v rámci svazku
4
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
22
Strana od-do
225-246
Kód UT WoS článku
000327364200001