A robust PPG-based heart rate monitor for fitness and ehealth applications
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989100%3A27240%2F18%3A10241623" target="_blank" >RIV/61989100:27240/18:10241623 - isvavai.cz</a>
Výsledek na webu
<a href="https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=8531082" target="_blank" >https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=8531082</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/HealthCom.2018.8531082" target="_blank" >10.1109/HealthCom.2018.8531082</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
A robust PPG-based heart rate monitor for fitness and ehealth applications
Popis výsledku v původním jazyce
This paper presents the design and implementation of a Heart Rate Monitor (HRM) based on photoplethysmography (PPG). This optical method is noninvasive and provides a convenient means of implementing a low-cost wearable HRM for eHealth applications. Our novel design consists of a light source that emits a modulated or unmodulated light signal through the finger tissue and measures the changes in the reflected light from the arterial blood. These reflections correspond well with the blood volume changes in synchrony with the heartbeat. Here we aim to provide a detailed description of how to design and implement an HRM device based on two approaches: 1) PPG using an unmodulated light source; 2) PPG using a modulated light source. Noise analysis in these devices enabled us to conceptualize and compare their performance and infer which one offers a better choice for optical heart rate monitoring. Based on the measurement results, the latter approach offers superior performance due to better noise cancellation and higher Signal-to-Noise Ratio (SNR) and is therefore robust against movement artefacts, power line noise, flicker noise of electronic components, as well as background environmental light interference. (C) 2018 IEEE.
Název v anglickém jazyce
A robust PPG-based heart rate monitor for fitness and ehealth applications
Popis výsledku anglicky
This paper presents the design and implementation of a Heart Rate Monitor (HRM) based on photoplethysmography (PPG). This optical method is noninvasive and provides a convenient means of implementing a low-cost wearable HRM for eHealth applications. Our novel design consists of a light source that emits a modulated or unmodulated light signal through the finger tissue and measures the changes in the reflected light from the arterial blood. These reflections correspond well with the blood volume changes in synchrony with the heartbeat. Here we aim to provide a detailed description of how to design and implement an HRM device based on two approaches: 1) PPG using an unmodulated light source; 2) PPG using a modulated light source. Noise analysis in these devices enabled us to conceptualize and compare their performance and infer which one offers a better choice for optical heart rate monitoring. Based on the measurement results, the latter approach offers superior performance due to better noise cancellation and higher Signal-to-Noise Ratio (SNR) and is therefore robust against movement artefacts, power line noise, flicker noise of electronic components, as well as background environmental light interference. (C) 2018 IEEE.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
<a href="/cs/project/EF16_019%2F0000867" target="_blank" >EF16_019/0000867: Centrum výzkumu pokročilých mechatronických systémů</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2018
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
2018 IEEE 20th International Conference on e-Health Networking, Applications and Services, Healthcom 2018