A comparative study of single-channel signal processing methods in fetal phonocardiography
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989100%3A27240%2F22%3A10250767" target="_blank" >RIV/61989100:27240/22:10250767 - isvavai.cz</a>
Výsledek na webu
<a href="https://journals.plos.org/plosone/article?id=10.1371/journal.pone.0269884" target="_blank" >https://journals.plos.org/plosone/article?id=10.1371/journal.pone.0269884</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1371/journal.pone.0269884" target="_blank" >10.1371/journal.pone.0269884</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
A comparative study of single-channel signal processing methods in fetal phonocardiography
Popis výsledku v původním jazyce
Fetal phonocardiography is a non-invasive, completely passive and low-cost method based on sensing acoustic signals from the maternal abdomen. However, different types of interference are sensed along with the desired fetal phonocardiography. This study focuses on the comparison of fetal phonocardiography filtering using eight algorithms: Savitzky-Golay filter, finite impulse response filter, adaptive wavelet transform, maximal overlap discrete wavelet transform, variational mode decomposition, empirical mode decomposition, ensemble empirical mode decomposition, and complete ensemble empirical mode decomposition with adaptive noise. The effectiveness of those methods was tested on four types of interference (maternal sounds, movement artifacts, Gaussian noise, and ambient noise) and eleven combinations of these disturbances. The dataset was created using two synthetic records r01 and r02, where the record r02 was loaded with higher levels of interference than the record r01. The evaluation was performed using the objective parameters such as accuracy of the detection of S1 and S2 sounds, signal-to-noise ratio improvement, and mean error of heart interval measurement. According to all parameters, the best results were achieved using the complete ensemble empirical mode decomposition with adaptive noise method with average values of accuracy = 91.53% in the detection of S1 and accuracy = 68.89% in the detection of S2. The average value of signal-to-noise ratio improvement achieved by complete ensemble empirical mode decomposition with adaptive noise method was 9.75 dB and the average value of the mean error of heart interval measurement was 3.27 ms.
Název v anglickém jazyce
A comparative study of single-channel signal processing methods in fetal phonocardiography
Popis výsledku anglicky
Fetal phonocardiography is a non-invasive, completely passive and low-cost method based on sensing acoustic signals from the maternal abdomen. However, different types of interference are sensed along with the desired fetal phonocardiography. This study focuses on the comparison of fetal phonocardiography filtering using eight algorithms: Savitzky-Golay filter, finite impulse response filter, adaptive wavelet transform, maximal overlap discrete wavelet transform, variational mode decomposition, empirical mode decomposition, ensemble empirical mode decomposition, and complete ensemble empirical mode decomposition with adaptive noise. The effectiveness of those methods was tested on four types of interference (maternal sounds, movement artifacts, Gaussian noise, and ambient noise) and eleven combinations of these disturbances. The dataset was created using two synthetic records r01 and r02, where the record r02 was loaded with higher levels of interference than the record r01. The evaluation was performed using the objective parameters such as accuracy of the detection of S1 and S2 sounds, signal-to-noise ratio improvement, and mean error of heart interval measurement. According to all parameters, the best results were achieved using the complete ensemble empirical mode decomposition with adaptive noise method with average values of accuracy = 91.53% in the detection of S1 and accuracy = 68.89% in the detection of S2. The average value of signal-to-noise ratio improvement achieved by complete ensemble empirical mode decomposition with adaptive noise method was 9.75 dB and the average value of the mean error of heart interval measurement was 3.27 ms.
Klasifikace
Druh
J<sub>SC</sub> - Článek v periodiku v databázi SCOPUS
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
<a href="/cs/project/EF16_019%2F0000867" target="_blank" >EF16_019/0000867: Centrum výzkumu pokročilých mechatronických systémů</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2022
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
PLoS One
ISSN
1932-6203
e-ISSN
1932-6203
Svazek periodika
17
Číslo periodika v rámci svazku
August 19, 2022
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
32
Strana od-do
1-32
Kód UT WoS článku
—
EID výsledku v databázi Scopus
2-s2.0-85136109400