Kombinace hromadění velkého objemu vzorku a dynamického pH spoje pro on-line prekoncentraci slabých elektrolytů v kapilární elektroforéze ve srovnání s izotachoforézou
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989592%3A15310%2F07%3A00004825" target="_blank" >RIV/61989592:15310/07:00004825 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Combination of large volume sample stacking and dynamic pH junction for on-line preconcentration of weak electrolytes by capillary electrophoresis in comparison with isotachophoretic techniques
Popis výsledku v původním jazyce
An on-line preconcentration capillary electrophoresis (CE) technique, which combines a large volume sample stacking with a dynamic pH junction technique, is introduced in this paper. This dynamic pH junction with co-electroosmotic migration is formed between sodium borate pH 9.5 and sodium phosphate pH 2.5 with 150 mM sodium dodecylsulfate (SDS). A full capillary based injection allows determination of weak acidic compounds at ppb concentration levels (achieved LOD for benzoic acid was 11 nmol L-1). Theproposed preconcentration method was compared with ITP/ITP (LOD 120 nmol L-1), ITP/CZE (LOD 740 nmol L-1) and a simple CZE method (LOD 23,330 nmol L-1). The analytical potential of this method was assessed with juice test samples.
Název v anglickém jazyce
Combination of large volume sample stacking and dynamic pH junction for on-line preconcentration of weak electrolytes by capillary electrophoresis in comparison with isotachophoretic techniques
Popis výsledku anglicky
An on-line preconcentration capillary electrophoresis (CE) technique, which combines a large volume sample stacking with a dynamic pH junction technique, is introduced in this paper. This dynamic pH junction with co-electroosmotic migration is formed between sodium borate pH 9.5 and sodium phosphate pH 2.5 with 150 mM sodium dodecylsulfate (SDS). A full capillary based injection allows determination of weak acidic compounds at ppb concentration levels (achieved LOD for benzoic acid was 11 nmol L-1). Theproposed preconcentration method was compared with ITP/ITP (LOD 120 nmol L-1), ITP/CZE (LOD 740 nmol L-1) and a simple CZE method (LOD 23,330 nmol L-1). The analytical potential of this method was assessed with juice test samples.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2007
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Chromatography A: Symposium Volumes
ISSN
0021-9673
e-ISSN
—
Svazek periodika
1155
Číslo periodika v rámci svazku
2
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
6
Strana od-do
193-198
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—