Using Chromium Modified Carbon Paste Electrode for Heavy Metal Ions Determination
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F62156489%3A43210%2F16%3A43910238" target="_blank" >RIV/62156489:43210/16:43910238 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00216305:26620/16:PU122777
Výsledek na webu
<a href="https://mnet.mendelu.cz/mendelnet2016/mnet_2016_full.pdf" target="_blank" >https://mnet.mendelu.cz/mendelnet2016/mnet_2016_full.pdf</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Using Chromium Modified Carbon Paste Electrode for Heavy Metal Ions Determination
Popis výsledku v původním jazyce
This paper describes the preparation and electrochemical application of chromium (III) oxide modified carbon paste electrode for simple and high sensitive simultaneous determination of zinc, cadmium, lead and copper ions. Square wave anodic stripping voltammetry which is very sensitive for metal ions was selected as electrochemical detection method. This method was optimized with respect to deposition time, deposition potential, frequency and amplitude. The detection limits were found to be 30.3 μg/L for Zn2+, 3 μg/L for Cd2+, 24.2 μg/L and 3 μg/L for Pb2+ and Cu2+ respectively. The potential for simultaneous detection of these heavy metal ions by the chromium modified carbon paste electrode was also demonstrated. Using this modified carbon paste is also possible to detect mercury ions, detection limit was found to be 18.2 μg/L.
Název v anglickém jazyce
Using Chromium Modified Carbon Paste Electrode for Heavy Metal Ions Determination
Popis výsledku anglicky
This paper describes the preparation and electrochemical application of chromium (III) oxide modified carbon paste electrode for simple and high sensitive simultaneous determination of zinc, cadmium, lead and copper ions. Square wave anodic stripping voltammetry which is very sensitive for metal ions was selected as electrochemical detection method. This method was optimized with respect to deposition time, deposition potential, frequency and amplitude. The detection limits were found to be 30.3 μg/L for Zn2+, 3 μg/L for Cd2+, 24.2 μg/L and 3 μg/L for Pb2+ and Cu2+ respectively. The potential for simultaneous detection of these heavy metal ions by the chromium modified carbon paste electrode was also demonstrated. Using this modified carbon paste is also possible to detect mercury ions, detection limit was found to be 18.2 μg/L.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
CG - Elektrochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2016
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
MendelNet 2016: Proceedings of International PhD Students Conference
ISBN
978-80-7509-443-8
ISSN
—
e-ISSN
—
Počet stran výsledku
5
Strana od-do
994-998
Název nakladatele
Mendelova univerzita v Brně
Místo vydání
Brno
Místo konání akce
Brno
Datum konání akce
9. 11. 2016
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
000392968500178