High pressure modification of the simple automated liquid chromatographic system for splitless nano column gradient separations
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081715%3A_____%2F13%3A00422993" target="_blank" >RIV/68081715:_____/13:00422993 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
High pressure modification of the simple automated liquid chromatographic system for splitless nano column gradient separations
Popis výsledku v původním jazyce
Our simple liquid chromatographic system for splitless gradient nanocolumn separations at backpressure up to 50 bars was modified to run analysis under the pressure up to 300 bars. Programmable syringe pump was equipped with a 250-L high-pressure syringeand mobile phase gradients were created and stored in relatively long capillary. Available linear gradients of volumes from 5 to 50 L traced by uracil are presented. Sample introduction was performed as separate step and real large volume of sample (tens of microliters) could be injected and preconcentrated on-column. Relative standard deviation of retention times and peak areas were received (RSD < 0.3 % and RSD < 5 % respectively).
Název v anglickém jazyce
High pressure modification of the simple automated liquid chromatographic system for splitless nano column gradient separations
Popis výsledku anglicky
Our simple liquid chromatographic system for splitless gradient nanocolumn separations at backpressure up to 50 bars was modified to run analysis under the pressure up to 300 bars. Programmable syringe pump was equipped with a 250-L high-pressure syringeand mobile phase gradients were created and stored in relatively long capillary. Available linear gradients of volumes from 5 to 50 L traced by uracil are presented. Sample introduction was performed as separate step and real large volume of sample (tens of microliters) could be injected and preconcentrated on-column. Relative standard deviation of retention times and peak areas were received (RSD < 0.3 % and RSD < 5 % respectively).
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/VG20112015021" target="_blank" >VG20112015021: Vývoj instrumentálních metodických postupů rychlé detekce a identifikace biologických agens v reálných vzorcích</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2013
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Chemické Listy. Roč. 107, Issue s3.
ISBN
—
ISSN
1213-7103
e-ISSN
—
Počet stran výsledku
3
Strana od-do
438-440
Název nakladatele
Česká společnost chemická
Místo vydání
Praha
Místo konání akce
Brno
Datum konání akce
12. 11. 2013
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000328730800063