Latest improvements in development of simple splitless gradient liquid chromatographic system for micro and nano columns
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26310%2F13%3APU104054" target="_blank" >RIV/00216305:26310/13:PU104054 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Latest improvements in development of simple splitless gradient liquid chromatographic system for micro and nano columns
Popis výsledku v původním jazyce
Our simple liquid chromatographic system for splitless gradient nanocolumn separations at backpressure up to 50 bars was modified to run analysis under the pressure up to 300 bars. Programmable syringe pump was equipped with a 250-ul high-pressure syringe and mobile phase gradients were created and stored in relatively long capillary. Available linear gradients of volumes from 5 to 50 ul traced by uracil are presented. Sample introduction was performed as separate step and real large volume of sample (tens of microliters) could be injected and pre-concentrated on-column. Relative standard deviation of retention times and peak areas were received (RSD < 0.3 % and RSD < 5 % respectively).
Název v anglickém jazyce
Latest improvements in development of simple splitless gradient liquid chromatographic system for micro and nano columns
Popis výsledku anglicky
Our simple liquid chromatographic system for splitless gradient nanocolumn separations at backpressure up to 50 bars was modified to run analysis under the pressure up to 300 bars. Programmable syringe pump was equipped with a 250-ul high-pressure syringe and mobile phase gradients were created and stored in relatively long capillary. Available linear gradients of volumes from 5 to 50 ul traced by uracil are presented. Sample introduction was performed as separate step and real large volume of sample (tens of microliters) could be injected and pre-concentrated on-column. Relative standard deviation of retention times and peak areas were received (RSD < 0.3 % and RSD < 5 % respectively).
Klasifikace
Druh
O - Ostatní výsledky
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2013
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů