Frequency method of sub-nanometer distance measurement using the optical resonator and tunable semiconductor laser.
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F02%3A12020137" target="_blank" >RIV/68081731:_____/02:12020137 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Frequency method of sub-nanometer distance measurement using the optical resonator and tunable semiconductor laser.
Popis výsledku v původním jazyce
Several methods of scale non-linearity measurements of Michelson interferometers have been presented but their uncertainty is comparable approximately with the total resolution of the interferometers. We developed and experimentally verified a new methodwhere order-times smaller uncertainty of the distance measurement is possible. The method takes advantage of a direct conversion of the relative changes of the optical path in measuring arm of the Michelson interferometer to relative changes of the resonant optical frequency of the Fabry-Pert resonator. Frequency changes of the resonator have been monitored by a beat-frequency comparson between tunable semiconductor laser and HeNeI.sub.2./sub. optical frequency standard at 633 nm.
Název v anglickém jazyce
Frequency method of sub-nanometer distance measurement using the optical resonator and tunable semiconductor laser.
Popis výsledku anglicky
Several methods of scale non-linearity measurements of Michelson interferometers have been presented but their uncertainty is comparable approximately with the total resolution of the interferometers. We developed and experimentally verified a new methodwhere order-times smaller uncertainty of the distance measurement is possible. The method takes advantage of a direct conversion of the relative changes of the optical path in measuring arm of the Michelson interferometer to relative changes of the resonant optical frequency of the Fabry-Pert resonator. Frequency changes of the resonator have been monitored by a beat-frequency comparson between tunable semiconductor laser and HeNeI.sub.2./sub. optical frequency standard at 633 nm.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
BH - Optika, masery a lasery
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2002
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of the general assembly of the international union of radio science.
ISBN
—
ISSN
—
e-ISSN
—
Počet stran výsledku
4
Strana od-do
"X1"-"X5"
Název nakladatele
N/A
Místo vydání
Maastricht
Místo konání akce
Maastricht [NL]
Datum konání akce
17. 8. 2002
Typ akce podle státní příslušnosti
WRD - Celosvětová akce