Collection of secondary electrons in scanning electron microscopes
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F09%3A00333617" target="_blank" >RIV/68081731:_____/09:00333617 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Collection of secondary electrons in scanning electron microscopes
Popis výsledku v původním jazyce
Collection of the secondary electrons in the scanning electron microscope was simulated and the results have been experimentally verified for two types of the objective lens and three detection systems. The aberration coefficients of both objective lenses as well as maximum axial magnetic fields in the specimen region are presented. Compared are a standard side-attached secondary electron detector, in which only weak electrostatic and nearly no magnetic field influence the signal trajectories in the specimen vicinity, and the side-attached (lower) and upper detectors in an immersion system with weak electrostatic but strong magnetic field penetrating towards the specimen. The collection efficiency was calculated for all three detection systems and several working distances. The ability of detectors to attract secondary electron trajectories for various initial azimuthal and polar angles was calculated, too.
Název v anglickém jazyce
Collection of secondary electrons in scanning electron microscopes
Popis výsledku anglicky
Collection of the secondary electrons in the scanning electron microscope was simulated and the results have been experimentally verified for two types of the objective lens and three detection systems. The aberration coefficients of both objective lenses as well as maximum axial magnetic fields in the specimen region are presented. Compared are a standard side-attached secondary electron detector, in which only weak electrostatic and nearly no magnetic field influence the signal trajectories in the specimen vicinity, and the side-attached (lower) and upper detectors in an immersion system with weak electrostatic but strong magnetic field penetrating towards the specimen. The collection efficiency was calculated for all three detection systems and several working distances. The ability of detectors to attract secondary electron trajectories for various initial azimuthal and polar angles was calculated, too.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2009
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Microscopy
ISSN
0022-2720
e-ISSN
—
Svazek periodika
236
Číslo periodika v rámci svazku
3
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
8
Strana od-do
—
Kód UT WoS článku
000271974200006
EID výsledku v databázi Scopus
—