Simulation-based optimisation of thermodynamic conditions in the ESEM for dynamical in-situ study of spherical polyelectrolyte complex particles in their native state
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F20%3A00525096" target="_blank" >RIV/68081731:_____/20:00525096 - isvavai.cz</a>
Výsledek na webu
<a href="https://www.sciencedirect.com/science/article/abs/pii/S0304399119303298" target="_blank" >https://www.sciencedirect.com/science/article/abs/pii/S0304399119303298</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.ultramic.2020.112954" target="_blank" >10.1016/j.ultramic.2020.112954</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Simulation-based optimisation of thermodynamic conditions in the ESEM for dynamical in-situ study of spherical polyelectrolyte complex particles in their native state
Popis výsledku v původním jazyce
We present a complex analysis and optimisation of dynamic conditions in the environmental scanning electron microscope (ESEM) to allow in-situ observation of extremely delicate wet bio-polymeric spherical particles in their native state. According to the results of gas flow and heat transfer simulations, we were able to develop an improved procedure leading to thermodynamic equilibrium between the sample and chamber environment. To quantify and hence minimise the extent of any sample deformation during specimen chamber pumping, a strength-stress analysis is used. Monte Carlo simulations of beam-gas,water, andsample interactions describe beam scattering, absorbed energy, interaction volume and the emission of signal electrons in the ESEM. Finally, we discuss sample damage as a result of drying and the production of beam-induced free radicals. Based on all experimental and simulation results we introduce a Delicate Sample Observation Strategy for the ESEM. We show how this strategy can be applied to the characterization of polyelectrolyte complex spherical particles containing immobilized recombinant cells E. coli overexpressing cyclohexanone monooxygenase, used as a model biocatalyst. We present the first native-state electron microscopy images of the viscous core of a halved polyelectrolyte complex capsule containing living cells.
Název v anglickém jazyce
Simulation-based optimisation of thermodynamic conditions in the ESEM for dynamical in-situ study of spherical polyelectrolyte complex particles in their native state
Popis výsledku anglicky
We present a complex analysis and optimisation of dynamic conditions in the environmental scanning electron microscope (ESEM) to allow in-situ observation of extremely delicate wet bio-polymeric spherical particles in their native state. According to the results of gas flow and heat transfer simulations, we were able to develop an improved procedure leading to thermodynamic equilibrium between the sample and chamber environment. To quantify and hence minimise the extent of any sample deformation during specimen chamber pumping, a strength-stress analysis is used. Monte Carlo simulations of beam-gas,water, andsample interactions describe beam scattering, absorbed energy, interaction volume and the emission of signal electrons in the ESEM. Finally, we discuss sample damage as a result of drying and the production of beam-induced free radicals. Based on all experimental and simulation results we introduce a Delicate Sample Observation Strategy for the ESEM. We show how this strategy can be applied to the characterization of polyelectrolyte complex spherical particles containing immobilized recombinant cells E. coli overexpressing cyclohexanone monooxygenase, used as a model biocatalyst. We present the first native-state electron microscopy images of the viscous core of a halved polyelectrolyte complex capsule containing living cells.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
20303 - Thermodynamics
Návaznosti výsledku
Projekt
<a href="/cs/project/GA19-03909S" target="_blank" >GA19-03909S: Pokročilé simulace interakcí elektronů s plynem pro vysoce účinnou detekci sekundárních elektronů při dynamických in-situ experimentech v EREM.</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2020
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Ultramicroscopy
ISSN
0304-3991
e-ISSN
—
Svazek periodika
211
Číslo periodika v rámci svazku
APR
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
15
Strana od-do
112954
Kód UT WoS článku
000522150300010
EID výsledku v databázi Scopus
2-s2.0-85078675621