ORGAN AND EFFECTIVE DOSES FROM VERIFICATION TECHNIQUES IN IMAGE-GUIDED RADIOTHERAPY (IGRT)
Popis výsledku v původním jazyce
The objective of this work was an evaluation and comparison of organ and effective doses from four verification techniques of Image-guided radiotherapy. These techniques were: megavoltage (MV) Cone beam CT (CBCT) scans, two orthogonal MV images, kilovoltage (kV) CBCT scans and two orthogonal kV images. Doses were evaluated for two treatment sites: pelvis and head and neck. Organ doses were measured in a male anthropomorphic Rando phantom using thermoluminiscent dosimeters (TLDs). Measurements were carried out on Siemens Artiste, Varian Clinac and Elekta Synergy linear accelerators. Radiotherapy departments can use presented dose data to estimate patient dose from their particular verification imaging technique.
Název v anglickém jazyce
ORGAN AND EFFECTIVE DOSES FROM VERIFICATION TECHNIQUES IN IMAGE-GUIDED RADIOTHERAPY (IGRT)
Popis výsledku anglicky
The objective of this work was an evaluation and comparison of organ and effective doses from four verification techniques of Image-guided radiotherapy. These techniques were: megavoltage (MV) Cone beam CT (CBCT) scans, two orthogonal MV images, kilovoltage (kV) CBCT scans and two orthogonal kV images. Doses were evaluated for two treatment sites: pelvis and head and neck. Organ doses were measured in a male anthropomorphic Rando phantom using thermoluminiscent dosimeters (TLDs). Measurements were carried out on Siemens Artiste, Varian Clinac and Elekta Synergy linear accelerators. Radiotherapy departments can use presented dose data to estimate patient dose from their particular verification imaging technique.
Klasifikace
Druh
A - Audiovizuální tvorba
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/JC_4%2F2009" target="_blank" >JC_4/2009: Stanovování a ověřování dávek při pokročilých radioterapeutických metodách (IMRT, IGRT).</a><br>
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
ISBN
—
Místo vydání
Bonn
Název nakladatele resp. objednatele
—
Verze
—
Identifikační číslo nosiče
—