Relation between left ventricular unloading during ECMO and drainage catheter size assessed by mathematical modeling
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21460%2F17%3A00315121" target="_blank" >RIV/68407700:21460/17:00315121 - isvavai.cz</a>
Výsledek na webu
<a href="https://ojs.cvut.cz/ojs/index.php/ap/article/download/4185/4405" target="_blank" >https://ojs.cvut.cz/ojs/index.php/ap/article/download/4185/4405</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.14311/AP.2017.57.0367" target="_blank" >10.14311/AP.2017.57.0367</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Relation between left ventricular unloading during ECMO and drainage catheter size assessed by mathematical modeling
Popis výsledku v původním jazyce
The flow-dependent left ventricle overload is a well-known complication of the veno-arterial extracorporeal membrane oxygenation in a severe cardiogenic shock, which leads to a distension of the left ventricle and, frequently, to a severe pulmonary edema. Recently, an unloading of the left ventricle using a catheter inserted to the left ventricle and connected to the extracorporeal membrane oxygenation circuit has been proposed. The computational method was used to simulate the blood flow in the extracorporeal membrane oxygenation system with a drainage catheter incorporated to the left ventricle and connected to the inflow part of the extracorporeal membrane oxygenation circuit by a Y-shaped connector. The whole system was modelled in Modelica modelling language. The impact of various catheter sizes (from 5Fr to 10Fr) and extracorporeal blood flow values (from 1 L/min to 5 L/min) were investigated. In our simulation model, the extracorporeal blood flow only modestly affected the value of volume that was withdrawn from the left ventricle by a catheter. Conversely, the size of the drainage catheter was the principal factor responsible for the achievement of the adequate left ventricle decompression. A 10 Fr drainage catheter, inserted into the left ventricle and connected to the venous part of the ECMO system, presents a promising solution to the unloading of the left ventricle during a extracorporeal membrane oxygenation.
Název v anglickém jazyce
Relation between left ventricular unloading during ECMO and drainage catheter size assessed by mathematical modeling
Popis výsledku anglicky
The flow-dependent left ventricle overload is a well-known complication of the veno-arterial extracorporeal membrane oxygenation in a severe cardiogenic shock, which leads to a distension of the left ventricle and, frequently, to a severe pulmonary edema. Recently, an unloading of the left ventricle using a catheter inserted to the left ventricle and connected to the extracorporeal membrane oxygenation circuit has been proposed. The computational method was used to simulate the blood flow in the extracorporeal membrane oxygenation system with a drainage catheter incorporated to the left ventricle and connected to the inflow part of the extracorporeal membrane oxygenation circuit by a Y-shaped connector. The whole system was modelled in Modelica modelling language. The impact of various catheter sizes (from 5Fr to 10Fr) and extracorporeal blood flow values (from 1 L/min to 5 L/min) were investigated. In our simulation model, the extracorporeal blood flow only modestly affected the value of volume that was withdrawn from the left ventricle by a catheter. Conversely, the size of the drainage catheter was the principal factor responsible for the achievement of the adequate left ventricle decompression. A 10 Fr drainage catheter, inserted into the left ventricle and connected to the venous part of the ECMO system, presents a promising solution to the unloading of the left ventricle during a extracorporeal membrane oxygenation.
Klasifikace
Druh
J<sub>SC</sub> - Článek v periodiku v databázi SCOPUS
CEP obor
—
OECD FORD obor
20601 - Medical engineering
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Acta Polytechnica
ISSN
1210-2709
e-ISSN
—
Svazek periodika
57
Číslo periodika v rámci svazku
5
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
6
Strana od-do
367-372
Kód UT WoS článku
000418522700008
EID výsledku v databázi Scopus
2-s2.0-85033376421