Effect of electrode gel application between patient’s skin and electrode belt on electrical impedance tomography of the thorax
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21460%2F19%3A00321935" target="_blank" >RIV/68407700:21460/19:00321935 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1007/978-981-10-9035-6" target="_blank" >http://dx.doi.org/10.1007/978-981-10-9035-6</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1007/978-981-10-9035-6" target="_blank" >10.1007/978-981-10-9035-6</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Effect of electrode gel application between patient’s skin and electrode belt on electrical impedance tomography of the thorax
Popis výsledku v původním jazyce
Electrical impedance tomography (EIT) is a non-invasive, radiation-free imaging method that can be used for bed-side monitoring of the distribution of ventilation in lungs. However, using EIT for lung monitoring in research and clinical practice is still limited by many methodical issues. In this case study, we analyzed the possible effect the application of electrode gel between the subject’s skin and the electrode belt can have on the EIT lung image. If the gel is wrongly applied, it can create a conductive connection between neighboring electrodes and distort the EIT signal. Our pilot study first minimized several unpredicted factors causing changes in EIT image and then confirmed that the conductive connection of the electrodes can lead to significant changes in the evaluated distribution of ventilation. Based on our preliminary findings, a greater study will follow with the aim of providing an official recommendation of how the gel should be applied to avoid possible errors that could lead to inaccurate conclusions about patients’ health status.
Název v anglickém jazyce
Effect of electrode gel application between patient’s skin and electrode belt on electrical impedance tomography of the thorax
Popis výsledku anglicky
Electrical impedance tomography (EIT) is a non-invasive, radiation-free imaging method that can be used for bed-side monitoring of the distribution of ventilation in lungs. However, using EIT for lung monitoring in research and clinical practice is still limited by many methodical issues. In this case study, we analyzed the possible effect the application of electrode gel between the subject’s skin and the electrode belt can have on the EIT lung image. If the gel is wrongly applied, it can create a conductive connection between neighboring electrodes and distort the EIT signal. Our pilot study first minimized several unpredicted factors causing changes in EIT image and then confirmed that the conductive connection of the electrodes can lead to significant changes in the evaluated distribution of ventilation. Based on our preliminary findings, a greater study will follow with the aim of providing an official recommendation of how the gel should be applied to avoid possible errors that could lead to inaccurate conclusions about patients’ health status.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20601 - Medical engineering
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2019
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
World Congress on Medical Physics and Biomedical Engineering 2018 (Vol. 1)
ISBN
978-981-10-9034-9
ISSN
1680-0737
e-ISSN
—
Počet stran výsledku
5
Strana od-do
81-85
Název nakladatele
Springer Nature Singapore Pte Ltd.
Místo vydání
—
Místo konání akce
Prague
Datum konání akce
3. 6. 2018
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—