Response of current photon personal dosemeters to new operational quantities
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F86652052%3A_____%2F20%3AN0000059" target="_blank" >RIV/86652052:_____/20:N0000059 - isvavai.cz</a>
Výsledek na webu
<a href="https://academic.oup.com/rpd/article-abstract/190/1/45/5850960" target="_blank" >https://academic.oup.com/rpd/article-abstract/190/1/45/5850960</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/rpd/ncaa078" target="_blank" >10.1093/rpd/ncaa078</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Response of current photon personal dosemeters to new operational quantities
Popis výsledku v původním jazyce
The ICRU proposed new operational quantities, which are defined in close relation to effective dose and absorbed dose. Most of the current personal dosemeters were designed to measure personal dose equivalents. Because of differences between the new and old quantities, the existing dosemeters may not be ideal for measurements of the new quantities, personal dose, Hp, and absorbed dose in local skin, Dp local skin. For photon radiation sources, we investigated relative responses of the current personal dosemeters with respect to the measurement of the new quantities. Introduction of the new quantities into practice will require some changes in practical radiation protection. A re-calibration of the current dosemeters will be essential for Hp measurements in higher photon energy region (> 50 keV). For lower photon energies, a re-design of specific evaluation algorithms will be necessary. For purposes of Dp local skin measurements the dosemeters do not require any fundamental changes.
Název v anglickém jazyce
Response of current photon personal dosemeters to new operational quantities
Popis výsledku anglicky
The ICRU proposed new operational quantities, which are defined in close relation to effective dose and absorbed dose. Most of the current personal dosemeters were designed to measure personal dose equivalents. Because of differences between the new and old quantities, the existing dosemeters may not be ideal for measurements of the new quantities, personal dose, Hp, and absorbed dose in local skin, Dp local skin. For photon radiation sources, we investigated relative responses of the current personal dosemeters with respect to the measurement of the new quantities. Introduction of the new quantities into practice will require some changes in practical radiation protection. A re-calibration of the current dosemeters will be essential for Hp measurements in higher photon energy region (> 50 keV). For lower photon energies, a re-design of specific evaluation algorithms will be necessary. For purposes of Dp local skin measurements the dosemeters do not require any fundamental changes.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10304 - Nuclear physics
Návaznosti výsledku
Projekt
<a href="/cs/project/VI20192022156" target="_blank" >VI20192022156: Dozimetrie pro radiační nehody a incidenty v kontextu nových operačních veličin pro externí záření</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2020
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radiation Protection Dosimetry
ISSN
0144-8420
e-ISSN
1742-3406
Svazek periodika
190
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
13
Strana od-do
45-57
Kód UT WoS článku
000581760800006
EID výsledku v databázi Scopus
2-s2.0-85089129753