Voltammetric Determination of Homovanillic Acid on a Disposable Electrochemical Measuring Cell System with Integrated Carbon Composite Film Electrodes
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11310%2F15%3A10320390" target="_blank" >RIV/00216208:11310/15:10320390 - isvavai.cz</a>
Výsledek na webu
<a href="http://www.bestservis.eu/upload/file/Sbornik_metody15.pdf" target="_blank" >http://www.bestservis.eu/upload/file/Sbornik_metody15.pdf</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Voltammetric Determination of Homovanillic Acid on a Disposable Electrochemical Measuring Cell System with Integrated Carbon Composite Film Electrodes
Popis výsledku v původním jazyce
Array of carbon composite film electrodes embedded in 96-well microtitration plate was applied for DPV determination of oxidizable tumour biomarker homovanillic acid (HVA). For the preparation of composite electrodes, graphitic conductive microparticlesand nonconductive polystyrene were used. For the measurement, a buffer of pH 2 was selected as the optimum medium. In this medium, concentration dependences were measured; calculated HVA quantification limit was 0.3 mu mol L-1 and detection limit was 0.1mu mol L-1. Sorption of HVA on employed working electrode was examined in order to verify, if the current is influenced by the time period between cell filling and measurement. The sorption was not observed and current was stable even after 10 minutes.
Název v anglickém jazyce
Voltammetric Determination of Homovanillic Acid on a Disposable Electrochemical Measuring Cell System with Integrated Carbon Composite Film Electrodes
Popis výsledku anglicky
Array of carbon composite film electrodes embedded in 96-well microtitration plate was applied for DPV determination of oxidizable tumour biomarker homovanillic acid (HVA). For the preparation of composite electrodes, graphitic conductive microparticlesand nonconductive polystyrene were used. For the measurement, a buffer of pH 2 was selected as the optimum medium. In this medium, concentration dependences were measured; calculated HVA quantification limit was 0.3 mu mol L-1 and detection limit was 0.1mu mol L-1. Sorption of HVA on employed working electrode was examined in order to verify, if the current is influenced by the time period between cell filling and measurement. The sorption was not observed and current was stable even after 10 minutes.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
CG - Elektrochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GBP206%2F12%2FG151" target="_blank" >GBP206/12/G151: Centrum nových přístupů k bioanalýze a molekulární diagnostice</a><br>
Návaznosti
S - Specificky vyzkum na vysokych skolach<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Sborník přednášek mezinárodní odborné konference XXXV. Moderní elektrochemické metody
ISBN
978-80-905221-3-8
ISSN
—
e-ISSN
—
Počet stran výsledku
5
Strana od-do
133-137
Název nakladatele
Srsenová Lenka - Best servis Ústí nad Labem
Místo vydání
Ústí nad Labem, Czech Republic
Místo konání akce
Jetřichovice, Czech Republic
Datum konání akce
18. 5. 2015
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
000358792900031