Electrochemical Immunosensor Coupled to Cyclone Air Sampler for Detection of Escherichia coli DH5 alpha in Bioaerosols
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14310%2F12%3A00059725" target="_blank" >RIV/00216224:14310/12:00059725 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1002/elan.201100448" target="_blank" >http://dx.doi.org/10.1002/elan.201100448</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1002/elan.201100448" target="_blank" >10.1002/elan.201100448</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Electrochemical Immunosensor Coupled to Cyclone Air Sampler for Detection of Escherichia coli DH5 alpha in Bioaerosols
Popis výsledku v původním jazyce
The portable detector ImmunoSMART was combined with the cyclone air sampler to detect the model bioagent E. coli DH5a in bioaerosols. The selected capture antibody was immobilized on gold electrodes, antibody-peroxidase conjugate served as a tracer. In buffer, the amperometric immunosensor measured 103 to 108 CFU/mL within 30 min. Detection in air was realized in a closed aerosol chamber. Automated dissemination of the cells, sampling and measurement allowed remote testing of the cyclone/immunodetector.The level of 150 CFU/L in air was clearly indicated (20 min analysis including 5 min sampling) and independently confirmed using the slit sampler and agar cultivation. The results seem promising for further extended tests in both aerosol chamber and field trial conditions.
Název v anglickém jazyce
Electrochemical Immunosensor Coupled to Cyclone Air Sampler for Detection of Escherichia coli DH5 alpha in Bioaerosols
Popis výsledku anglicky
The portable detector ImmunoSMART was combined with the cyclone air sampler to detect the model bioagent E. coli DH5a in bioaerosols. The selected capture antibody was immobilized on gold electrodes, antibody-peroxidase conjugate served as a tracer. In buffer, the amperometric immunosensor measured 103 to 108 CFU/mL within 30 min. Detection in air was realized in a closed aerosol chamber. Automated dissemination of the cells, sampling and measurement allowed remote testing of the cyclone/immunodetector.The level of 150 CFU/L in air was clearly indicated (20 min analysis including 5 min sampling) and independently confirmed using the slit sampler and agar cultivation. The results seem promising for further extended tests in both aerosol chamber and field trial conditions.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CE - Biochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Electroanalysis
ISSN
1040-0397
e-ISSN
—
Svazek periodika
24
Číslo periodika v rámci svazku
3
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
8
Strana od-do
539-546
Kód UT WoS článku
000300931000011
EID výsledku v databázi Scopus
—