Voltammetric determination of leucovorin in pharmaceutical samples
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216275%3A25310%2F17%3A39910975" target="_blank" >RIV/00216275:25310/17:39910975 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Voltammetric determination of leucovorin in pharmaceutical samples
Popis výsledku v původním jazyce
Voltammetric behavior of leucovorin has been investigated for the first time using boron-doped diamond electrode (BDDE) and the method for its determination was developed. Leucovorin provides two well-developed oxidation peaks at about +900 and +1450 mV (pH 5) on BDDE. Britton-Robinson buffer (pH 3.0) was chosen as an optimal supporting electrolyte for its determination using differential pulse voltammetry (DPV) with optimized parameters. Limit of detection 1.5×10−8 mol L−1 was reached for leucovorin determination and linear dynamic range 1.5×10−7-2.5×10−5 M was obtained. Relative standard deviation of repeated measurements (RSDM(11) = 0.7 %) and relative standard deviations of repeated determinations (RSDD(5) < 3.0 %) were calculated to confirm very good results. Applicability of the proposed method was verified by an analysis of a pharmaceutical preparation.
Název v anglickém jazyce
Voltammetric determination of leucovorin in pharmaceutical samples
Popis výsledku anglicky
Voltammetric behavior of leucovorin has been investigated for the first time using boron-doped diamond electrode (BDDE) and the method for its determination was developed. Leucovorin provides two well-developed oxidation peaks at about +900 and +1450 mV (pH 5) on BDDE. Britton-Robinson buffer (pH 3.0) was chosen as an optimal supporting electrolyte for its determination using differential pulse voltammetry (DPV) with optimized parameters. Limit of detection 1.5×10−8 mol L−1 was reached for leucovorin determination and linear dynamic range 1.5×10−7-2.5×10−5 M was obtained. Relative standard deviation of repeated measurements (RSDM(11) = 0.7 %) and relative standard deviations of repeated determinations (RSDD(5) < 3.0 %) were calculated to confirm very good results. Applicability of the proposed method was verified by an analysis of a pharmaceutical preparation.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
10406 - Analytical chemistry
Návaznosti výsledku
Projekt
<a href="/cs/project/GA17-03868S" target="_blank" >GA17-03868S: Nové metody elektrochemického sledování biologicky aktivních organických látek v environmentálních, biologických a potravinových matricích</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
13th ISC Modern Analytical Chemistry
ISBN
978-80-7444-052-6
ISSN
—
e-ISSN
neuvedeno
Počet stran výsledku
6
Strana od-do
210-215
Název nakladatele
Univerzita Karlova. Přírodovědecká fakulta
Místo vydání
Praha
Místo konání akce
Praha
Datum konání akce
20. 9. 2017
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
000435216500038