Some notes to the design and implementation of the device for cord implants tuning
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26210%2F09%3APU83967" target="_blank" >RIV/00216305:26210/09:PU83967 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Some notes to the design and implementation of the device for cord implants tuning
Popis výsledku v původním jazyce
This contribution deals with a position control analysis of the biomechanical testing device based on the Stewart Platform. The constructed mechanism represents a six-degree of freedom positioning manipulator. Its closed kinematic chain and parallel linkage structure give it great rigidity and large load-to-weight ratio. That makes it suitable for testing of the backbone segments and hip joints. The mechanism constructed on FME BUT contains two plates (base and platform) which are connected with six linear mechanical actuators to the each other. The device control task is based on inverse position kinematics. A model of the mechanism inverse kinematics was built for the control purposes in MATLAB software. Finally the controller algorithm was designedand implemented into LabVIEW environment which is suitable for a real-time control.
Název v anglickém jazyce
Some notes to the design and implementation of the device for cord implants tuning
Popis výsledku anglicky
This contribution deals with a position control analysis of the biomechanical testing device based on the Stewart Platform. The constructed mechanism represents a six-degree of freedom positioning manipulator. Its closed kinematic chain and parallel linkage structure give it great rigidity and large load-to-weight ratio. That makes it suitable for testing of the backbone segments and hip joints. The mechanism constructed on FME BUT contains two plates (base and platform) which are connected with six linear mechanical actuators to the each other. The device control task is based on inverse position kinematics. A model of the mechanism inverse kinematics was built for the control purposes in MATLAB software. Finally the controller algorithm was designedand implemented into LabVIEW environment which is suitable for a real-time control.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JR - Ostatní strojírenství
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2009
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Recent advances in Mechatronics 2008 - 2009
ISBN
978-3-642-05021-3
ISSN
—
e-ISSN
—
Počet stran výsledku
6
Strana od-do
—
Název nakladatele
Springer-Verlag Berlin Heidelberg
Místo vydání
Berlin
Místo konání akce
Luhačovice
Datum konání akce
18. 11. 2009
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—