Design and performance of a flow sensor CoroQuant used with emergency lung ventilator CoroVent during COVID-19 pandemic
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21460%2F22%3A00358672" target="_blank" >RIV/68407700:21460/22:00358672 - isvavai.cz</a>
Výsledek na webu
<a href="https://doi.org/10.1016/j.measen.2022.100383" target="_blank" >https://doi.org/10.1016/j.measen.2022.100383</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.measen.2022.100383" target="_blank" >10.1016/j.measen.2022.100383</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Design and performance of a flow sensor CoroQuant used with emergency lung ventilator CoroVent during COVID-19 pandemic
Popis výsledku v původním jazyce
At the time of COVID-19 pandemic onset in spring 2020 a project CoroVent was initiated with the aim to design and produce emergency lung ventilators and distribute them to hospitals. No flow and tidal volume sensors were available for the project. The lack of tidal volume sensors was a consequence of the rapidly increased demand for mechanical lung ventilators and their consumables. The aim of the study was to develop a special flow sensor CoroQuant for the CoroVent ventilators. The sensor based on pneumotachographic principle, manufactured by the plastic injection moulding of polypropylene, meets the requirements for precision of tidal volume measurement defined by international standard ISO 80601-2-12 for mechanical lung ventilators. CoroVent ventilators with CoroQuant sensors were distributed to 27 hospitals in the Czech Republic for free upon the requests from the medical facilities and started to be clinically used, thus preventing lack of lung ventilators in hospitals in the Czech Republic.
Název v anglickém jazyce
Design and performance of a flow sensor CoroQuant used with emergency lung ventilator CoroVent during COVID-19 pandemic
Popis výsledku anglicky
At the time of COVID-19 pandemic onset in spring 2020 a project CoroVent was initiated with the aim to design and produce emergency lung ventilators and distribute them to hospitals. No flow and tidal volume sensors were available for the project. The lack of tidal volume sensors was a consequence of the rapidly increased demand for mechanical lung ventilators and their consumables. The aim of the study was to develop a special flow sensor CoroQuant for the CoroVent ventilators. The sensor based on pneumotachographic principle, manufactured by the plastic injection moulding of polypropylene, meets the requirements for precision of tidal volume measurement defined by international standard ISO 80601-2-12 for mechanical lung ventilators. CoroVent ventilators with CoroQuant sensors were distributed to 27 hospitals in the Czech Republic for free upon the requests from the medical facilities and started to be clinically used, thus preventing lack of lung ventilators in hospitals in the Czech Republic.
Klasifikace
Druh
J<sub>SC</sub> - Článek v periodiku v databázi SCOPUS
CEP obor
—
OECD FORD obor
20601 - Medical engineering
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2022
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Measurement: Sensors
ISSN
2665-9174
e-ISSN
—
Svazek periodika
22
Číslo periodika v rámci svazku
100383
Stát vydavatele periodika
CH - Švýcarská konfederace
Počet stran výsledku
5
Strana od-do
1-5
Kód UT WoS článku
—
EID výsledku v databázi Scopus
2-s2.0-85132874428